Functions
100
© SICK AG • Subject to change without notice. • 8017035/ZMO9/2017-06-08
Tips: •
In order to perform the diameter measurement, the profile must cross
the left and right sides of the area (the thick-bordered frame). If the
profile does not cross both sides, the measured result will indicate that
measurement is not possible.
• You can perform a more precise measurement by ensuring that the top
of the arc is positioned in the center of the area and by measuring over
a wide range.
Inflect
When the curvature of the point of the profile meets or exceeds the level
set, output its position. When it can’t detect the point meets the level,
output result shows “----”.
Tips: •
Inflection graph is not available.
7.7
Calculation Settings
Sets the processing to perform in relation to the measured values
of each area.
7.7.1
Average
Averaging is performed the number of times specified over the measured
results of the areas.
The more times specified for processing, the more stable the result,
but the response speed is slower.
[Setting target:] Areas 1 to 4
Tips: •
If you have used “Smoothing” as described in “7.5.1 Profile Extraction
Settings,” the response speed will not be changed.
7.7.2
Hold
You can use hold mode to acquire the maximum or minimum value in the
specified length of time.
By detecting the maximum or minimum value while the workpiece passes
by, you can easily check for problems such as the occurrence of abnormal
values.
[Setting target:] Areas 1 to 4 and calculations 1 and 2
None
The hold operation is not performed. (Hold input will be ignored.)
Sample
The measured values during the hold input period are maintained.
Peak
The maximum value during the hold input period is output.
Bottom
The minimum value during the hold input period is output.