Congratulations, you have acquired a

 SIBELMED 

spirometer, a leading brand in the 

industry.
Please read the

 USER MANUAL

 in order to maximize its using.

The

 

DATOSPIR

 

touch

 

spirometer is a compact device based in different types of 

transducers:  Fleisch,  Turbine or disposable Lilly, a wide color touch screen, a built-in 
temperature sensor and a thermal printer. An internal database is used to store the 
tests and an external printer can be connected via USB or Bluetooth.

SETUP

When placing the spirometer make sure that the following conditions are meet:
- Mains connection: 100 to 240V AC.
- Operating temperature: 5 to 40º C
- Relative humidity: < 85 % (without condensation).
- Pressure: 850 to 1060 hPa. (638 to 795 mmHg / 1500 to 0 m aprox.)
- Keep away from splashing water or other liquids.

Note:  

Read the user’s manual for a safe use of the device.

Use  mouthpieces  certified  acording  to  Directive  93/42/EEC  to 
ensure the patient safety and the quality of spirometry.

CALIBRATION

(C) 

Making a calibration:

1.

 

Press 

 in the main menu.

2.

 

Press 

 in the setup menu.

3.

 

Press 

 in the calibration menu.

Repeat points 2 and 3 of the previous section

 

(B).

5.

 

Press 

 to save the calibration. 

4.

 

Once the calibration cycles are completed, the device 

should display the “CALIBRATED !!!” message.

The difference in this mode is that the device not only checks the calibration, 
but also stores the new calibration factors.