- 35 -
Entry is rejected if the upper limit value is
larger than the lower limit value.
Entry is rejected if the upper limit value is
larger than the lower limit value.
Setting the Error
In the menu screen, press TRBL SET to display the error setting screen as in the figure
below. Press the respective numerical regions to proceed to the numerical value entry screen.
Set the respective threshold values. Press ESC to return to the menu screen.
The respective numerical value entry screens are as follows. Press the numerical keys to
enter the desired numerical values.
Press
to confirm the numerical values as the setting values and return to the error setting
screen.
Press ESC to cancel entry and return to the error setting screen.
Press CLR to set the numerical value to "0."
Press BS to delete the last figure entered.
Enter a flow rate accuracy threshold value
between 5 % and 30 %.
The factory default is
±
30 %.
Enter a suction pressure threshold value
between -250 hPa and 0.0 hPa.
The factory default upper limit is 0.0 hPa, and
the lower limit is -250 hPa.
Enter an atmospheric temperature threshold
value between -20.0
°
C and 50.0
°
C.
The factory default upper limit is 50.0
°
C, and
the lower limit is -20.0
°
C.
ESC key
Suction pressure lower
limit threshold
Suction pressure upper
limit threshold
Atmospheric temperature
upper limit threshold
Atmospheric temperature
lower limit threshold
Flow rate accuracy threshold