- 15 -
Note however that suction capacity will be reduced when this product operates at a low
pressure site.
Capability of This Product
This product can perform a variety of dust collection procedures using various types of
optional equipment.
It is capable of simultaneous measurement of atmospheric pressure and temperature, and
actual flow rate control, meaning that inertial impact particle size selector, which is sensitive to
flow velocity, can be attached. In addition, it is capable of operating at high altitude sites and
other places with differing air pressure.
Tube
Through pipe:
HV-RW for dust collection standard accessory
Shuttle tube:
HV-RW for dioxins standard accessory. It contains a urethane holder
(glass tube), containing a 90 mm diameter polyurethane foam that is
used to perform dioxin sampling.
Filter
Filters:
8"
×
10" filters can be mounted.
Particle size selector: A special inertial impact particle size selector can be attached to
perform PM2.5 and other size classification.
Andersen sampler:
An AH-600 Andersen sampler can be attached to perform size
classification.
Orifice flow meter:
A special orifice flow meter can be attached to perform flow rate
calibrations.
Others
We accept special orders for additional connecting parts and special measurements. Contact
the distributor or your Sibata representative.
Product Limitations
This product is a suction based sampler. Note that it is not capable of the following:
Utilization of exhaust
Intake of anything (chemicals or gasses) except air
Intake of water or other liquids