                                   Tomsk Scientific Industrial Introduction Company “SIAM” LLC 

 
 

 

 

 
 

 

WELL SOUNDERS 

«DU-1». 

OPERATING MANUAL 

 

 
 

Explosion-proof design,

 

type of explosion protection – intrinsically safe circuit

 

1Ex ib IIB T3 Gb Х explosion proof mark

 

 

 
 
 
 
 
 
 
 
 
 
 
 
 

Russia 

Summary of Contents for DU-1

Page 1: ...entific Industrial Introduction Company SIAM LLC WELL SOUNDERS DU 1 OPERATING MANUAL Explosion proof design type of explosion protection intrinsically safe circuit 1Ex ib IIB T3 Gb explosion proof mar...

Page 2: ...Well sounders DU 1...

Page 3: ...OFNESS OF THE PRODUCT 13 7 EQUIPMENT INSTALLATION AND DEINSTALLATION ORDER 16 7 1 Preparation of Test Objects for Echometry 16 7 2 Well sounder Installation 16 7 3 Well sounder Removal 17 8 BATTERY CH...

Page 4: ...32 11 2 6 Replacement of the Valve Sealing Ring 33 12 DEVICE STORAGE AND TRANSPORTATION 34 APPENDIX 1 Control of Well Fluid Level in Complicated Conditions 35 APPENDIX 2 Setting Sound Speed Work Shee...

Page 5: ...range from 40 C to 50 C The device design is comply with the assigned explosion proof mark with the requirements of GOST IEC 60079 14 2013 and other regulatory documents governing the use of electric...

Page 6: ...31610 2011 and have 1Ex ib IIB T3 Gb explosion proof mark DU 1 well sounder is designed for indoor and outdoor installation in hazardous areas where atmospheric explosives of IIA IIB categories and T1...

Page 7: ...r consists of 1 the case 2 thread connecting coupling 3 two shackles 4 the rotary retrievable outlet valve 5 the ON OFF button 6 the Power LED 7 the Charge LED 8 the plug for charging the in built acc...

Page 8: ...e maintains its operability at ambient temperatures from minus 40 C to 50 C 2 3 The device operates autonomously and is powered by a special internal battery type EEMB LP103450LC PCM LD with 3 6 V or...

Page 9: ...ing to the set shown in Table 2 Table 2 Title Q ty 1 Well sounder DU 1 1 2 Certificate 1 3 Operating manual 1 4 DB Siam 2 5 software user s guide 1 5 CD with the software 1 6 Package 1 7 Conformity ce...

Page 10: ...rmed After the acoustic impulse is generated the rubber sphere should be kept in the position which it took immediately after the blow until the value of the measured level pops up on the screen The b...

Page 11: ...y approved by the RTSA 5 2 Requirements to Personnel Well tests are to be carried out by individuals above 18 who have secondary education and who are medically cleared for work Personnel are to take...

Page 12: ...tapers chokes hoppers and similar units The elbow of the offset should have pipe tapered thread 60 State Standard 633 80 and be located at the height 0 2 1 8 meters above the ground If the height exc...

Page 13: ...ces The minimum width of conductors on the PCB is 0 2 mm copper thickness is not less than 18 m Thus the electrical circuit coming out of the battery compartment of the device is intrinsically safe Th...

Page 14: ...hall be replaced by new ones supplied by the manufacturer 6 The device s operability is maintained at ambient temperature from minus 40 to plus 50 The maximum temperature of overheating of components...

Page 15: ...osion proofness during operation Check that the cover of the battery compartment and the compartment itself including the threaded connection are free from mechanical damage and corrosion In case of c...

Page 16: ...nnecting coupling of the sensor If the thread segment is worn out or significantly corroded or damaged well sounder installation is prohibited Open the valve on the offset for a short period of time 1...

Page 17: ...on the terminal indicator In order to avoid leaks in thread connections use 402 dope 38 101 330 73 or 0 1 20 sealing band 6 05 1388 86 7 3 Well sounder Removal Switch off the device Close the valve of...

Page 18: ...via an appropriate cable to a network or car adapter or USB port of your computer The charging time of a fully discharged battery is approximately 7 hours The voltage of a fully charged battery is 4 1...

Page 19: ...mputer you must connect the sensor using Windows Add Bluetooth device The procedure for connecting may vary depending on the OS version The sensor must be turned on and ready for connection The sign o...

Page 20: ...ck the right mouse button in the appeared context menu select item Open parameters see Figure 1 in the appeared Bluetooth Options window on the COM Ports tab see which outgoing port is assigned to the...

Page 21: ...ers DU 1 Working with devices in DB SIAM To work with devices in the DB SIAM software start the program switch to the Settings tab press the Sensors button see Fig 3 the Sensors dialog box will appear...

Page 22: ...blems a corresponding warning message appears This happens if the COM port is incorrectly specified or in case of the sensor is switched off If the error occurs because the sensor has been switched of...

Page 23: ...Well sounders DU 1 If the connection is successful the sensor control panel appears For each sensor type the control panel will have a different look...

Page 24: ...e sensor and allows the operator to save the study into the database at the moment when he is sure that he has correctly specified the well details and other study parameters If the operator has not s...

Page 25: ...ed to specify what sound speed to use for level detection You can use one of the tables of correction tables of dependence of sound speed in annular space on pressure in annular space or set the value...

Page 26: ...dy to start state and press the Start button 6 The Waiting for burst state would succeed the state Noise Measurement 7 Run the burst 8 The device will switch to the state of recording the echogram the...

Page 27: ...e 10 2 Multi purpose Valve The multi purpose valve is supplied separately and is to be used in SUDOS well sounders and SIAM manufactured well sounders for generating an acoustic impulse by injecting g...

Page 28: ...that allows to connect the conversion kit to the well sounder receiver The cylindrical receiver from GBO 2 kit is screwed on the outlet of the automatic valve unit or on the specialized manual valve o...

Page 29: ...12 months Cleaning of the tapered connecting thread the acoustic sensor the well sounder and the case Maintenance of the exhaust valve Cleaning of accumulator charging socket Performance monitoring 1...

Page 30: ...or When cleaning a very dirty thread application of metal tools an awl a screwdriver etc is allowed but do not apply an excessive force not to damage the thread It is strictly forbidden to use a metal...

Page 31: ...nents are given in the Figure 1 guide bush 2 tappet 3 retaining ring 4 spring 5 sealing ring 6 nut 7 cone 8 housing Washing is done with diesel fuel or kerosene using a brush and a rag Before washing...

Page 32: ...f valve housing Take out the guide bush 1 spring 4 tappet 2 in combination with cone 7 and nut 6 from valve housing 8 Then replace the cone 7 by unscrewing the nut 6 When necessary clean the contact c...

Page 33: ...with CIATIM GOST 6267 80 lubricant Then the valve should be put in place After a final assembly of the well sounder it is necessary to carry out the integrity and strength tests First install the wel...

Page 34: ...less than 80 Product transportation is allowed in a standard bag by any type of transport at the temperature from 50 to 50 Severe shock and vibration are to be avoided during transportation When stor...

Page 35: ...f the original signal and usually difficult to read In the annular space a level of interference is very high what is caused by an excessive vibration of the tubing hanger and casing ESP vibration val...

Page 36: ...of high interference are acoustic noises on the part of flowline In this case it is recommended to shut a casing valve of the flowline tree during measurement if it is not in conflict with the safety...

Page 37: ...in one field range The device will show the level based on the preset sound speed General use sheets are recorded in the read only memory of the device obtained from averaged data for respective regio...

Page 38: ...344 6 1 361 10 9 378 0 8 328 3 0 345 6 3 362 11 2 379 1 0 329 3 1 346 6 6 363 11 5 380 1 1 330 3 3 347 6 9 364 11 9 381 1 2 331 3 5 348 7 1 365 12 3 382 1 3 332 3 7 349 7 5 366 12 8 383 1 5 333 3 9 3...

Page 39: ...speed m s 0 7 310 2 8 327 6 3 340 13 5 351 0 8 311 3 0 329 6 6 341 14 0 352 1 0 313 3 1 329 6 9 341 15 0 353 1 1 314 3 3 330 7 1 342 17 0 354 1 2 315 3 5 331 7 5 343 20 0 356 1 3 316 3 7 332 7 7 343...

Page 40: ...E mail siamoil siamoil ru Web address http www siamoil ru For repair and maintenance of products manufactured by TSIIC SIAM LLC contact the service centers of the company 1 Service center in Neftejug...

Page 41: ...Well sounders DU 1 4 Service center in Tomsk 3 Belaya Str Tomsk 634003 Russian Federation tel 3822 90 00 08 20 02...

Page 42: ...Well sounders DU 1 REVISION RECORD SHEET Revi sion Numbers of sheets pages Total sheets pages of document Ref of docu ment Incoming of supporting document Signa ture Date Revised Replaced New Annulled...