Shodex IC SI-52 4E Operation Manual Download Page 1

ver.211025E

 

SHOWA DENKO K.K. (https://www.shodex.com/)

 

Operation Manual 

Shodex

 IC SI-52 4E 

 

Please read this manual carefully to achieve the best and consistent column performance for a long time

 

 
 

Important Handling Instructions 

 

 Caution! 

 

 

· Please consult the Safety Data Sheet (SDS) of reagents and solvents used with the column 

and understand their proper handling methods to prevent potential health hazards or death 
from occurring. 

 

· Please wear appropriate personal protective equipment such as lab goggles and gloves 

when handling organic solvents and acid and alkaline reagents. Avoid any direct physical 
contact to prevent chemical injuries.   

 

Before Using the Column 

(1) Please visually inspect the package and outside of the column for any damage.   

 

(2) Please check if product name and serial number written on the column package, column name tag, and     

enclosed Certificate of Analysis (CoA) are matching and correct.   

 

1. Introduction 

Thank you for purchasing the Shodex

 product. Shodex

 IC SI-52 4E is a suppressor type ion chromatography 

column for analysis of anions, using sodium carbonate eluents. The column is suitable for the 

simultaneous 

analysis of oxyhalides including chlorite, chlorite, and bromate, and standard inorganic anions including fluoride, 
chloride, nitrite, bromide, nitrate, phosphate, and sulfate. 
 

2. Column Components   
 
 

 
 
 

               

 

 

 
 
 
 
 
 
3. Column Specifications 

Product 

Code 

Product Name 

Column Size (mm) 

Particle Size   

(

μ

m) 

Theoretical Plate 

Number 

(Per Column) 

I.D. Length 

F6995260 

IC SI-52 4E 

4.0 250 

 14,000 

F6709626 

IC SI-92G 

4.6 

 

10 9 (Guard 

Column) 

 

Base Material 

: Spherical porous particles of polyvinyl alcohol modified with quaternary ammonium group

 

Column Housing  : PEEK   
Screw Type 

: Internally-threaded type No.10-32 UNF 

Shipping Solvent : 3.6 mM Sodium Carbonate Solution 
 

 

 

Summary of Contents for IC SI-52 4E

Page 1: ...and serial number written on the column package column name tag and enclosed Certificate of Analysis CoA are matching and correct 1 Introduction Thank you for purchasing the Shodex product Shodex IC SI 52 4E is a suppressor type ion chromatography column for analysis of anions using sodium carbonate eluents The column is suitable for the simultaneous analysis of oxyhalides including chlorite chlor...

Page 2: ...nated Always use freshly prepared solvents Solvents stored for a long time may have changed their compositions and may influence elution patterns and or damage the column Note Use of on line degasser is recommended 6 Sample Preparation 1 Filter the sample solution using disposable 0 45 μm filter to prevent the column from clogging or deteriorating 2 Recommended sample injection volume is less than...

Page 3: ...topped while the eluent inside the column was still hot as the eluent temperature decreases its volume also decreases This may result in creating an empty space in the column and deteriorates the column Note It is recommended to set the pump limiter to avoid exceeding the maximum pressure 7 3 Solvent Exchange When replacing the solvent start the system at less than 0 5 mL min Recommended solvent v...

Page 4: ...he column Do not suspend the column cleaning but complete all procedures until the last step to replace it with the eluent 8 Column Storage Remove the column from the system after replacing the in column solvent with the shipping solvent Securely tighten the end caps and store the column at a location with stable temperature a cool and dark space is recommended Refer to section 7 3 Solvent Exchang...

Page 5: ...ct on the column Do not drop or hit the column on a hard surface Refer to the Shodex website https www shodex com for product details and their applications For additional assistance contact the distributor from whom you purchased the column or contact your regional Shodex support office https www shodex com en support_office list ...

Page 6: ...anyone other than person or firm designated by Showa Denko K K 3 3 If the Shodex product is disposed of 3 4 If the Shodex product is resold by the user without giving prior written notice to Showa Denko K K 3 5 If the performance of the Shodex product is not conform to the specification of the attached CERTIFICATE OF ANALYSIS due to any of the reasons below a Computer virus b Impurities contained ...