Shodex Asahipak ODP-40 4D Operation Manual Download Page 1

Ver.140313

 

 

 

Showa Denko K.K.    (http://www.shodex.com/) 

Operation Manual 

Shodex

TM

 Asahipak

TM

 ODP-40 4D, 4E 

 

(Please read this manual carefully before using the column to keep its good performance and life.) 

 

1.Introduction 

The Shodex Asahipak ODP-40 4D and ODP-40 4E are columns for reversed phase chromatography. 

These columns are packed with spherical porous polymer having octadecyl (C18), and perform similar to 
general ODS columns. These columns are compatible with not only water (H

2

O) or polar organic solvents, but 

also the wide pH range between 2 to 13. 
 

2. Instructions in handling

 <Important> 

  Caution! 

    *  Take  notice  of  keeping  instructions  about  the  solvents  and  the  reagents  used  with  the 

column not to occur problems related to losing your health or leaking. 

  Attention! 

  * Use the column within the regular range of flow rate, pressure and temperature. There is a 

danger of deteriorating the performance when it is handled beyond the permissible range 
even for a short time.    See the clause 

“Usable Conditions” about the permissible range. 

 

3. Specifications 

Product code  Product name 

Size (mm) 

Particle size 

Plate Number 

(TP/per column) 

ID 

Length 

(μm) 

F7621001 

ODP-40 4D 

4.6 

150 

11,000 

F7621002 

ODP-40 4E 

4.6 

250 

17,000 

F6710022 

ODP-50G 4A 

4.6 

  10 

(Guard column) 

Packing material 

: Spherical porous particles of polyvinyl alcohol with chemical bonds of C18 groups   

Carbon contents 

: 17% 

Column material 

: Stainless steel (SUS 316) 

Screw type 

: Unified Thread Standard No. 10-32 UNF 

Shipping solvent 

: H

2

O/CH

3

CN = 35/65 vol % 

 

4. Usable Conditions 

Product name 

Flow rate

mL/min

 

Max pressure 

pH 

range 

Temperature

 

range 

Standard 

Max 

(MPa) 

 

ODP-40 4D 

0.3

0.6 

0.8 

13 

2

13 

4

60

 

ODP-40 4E 

0.3

0.6 

0.8 

ODP-50G 4A 

0.3

0.6 

0.8 

13 

    Apply lower flow rate than standard for solvent replacement. 
 

5. Solvents 

1)  Filtrate  the  eluent  with  a  membrane 

filter  (0.45μm)  to  prevent  chromatogram  noise  and  column 

performance deterioration by small particles or unsolved materials.. 

2) It is possible to use pure water, methanol and acetonitrile 100% each, and any proportion of these solvents.   

Acetonitrile  is  recommended  in  case  of  low  resolution  of  aromatic  substances  with  methanol  containing 
eluent. 

3) It is possible to use buffer agents like phosphoric salt, acetate salt and Tris-HCL. And neutral salts such as 

NaCl, KCl, Na

2

SO

4

 and acids such as trifluoroacetic acid are also usable. Check no precipitation of salts and 

solvent pH, and filtrate the eluent with a membrane filter 

(0.45μm) 

 

6. Sample preparation 

1) Filtrate the sample solution with a disposable 

filter (0.45μm) to prevent column performance deterioration 

by small particles or unsolved materials. 

2)  Make  sample  using  eluent  as  diluted  solution.  Starting  eluent  is  recommended  for  making  sample  for 

gradient analysis. 

 

7. Column Cleaning 

    Elution  characteristics  of  a  column  may  change  considerably  after  long,  repeated  usage,  due  to  the 

accumulation  of  sample  microadsorbents  on  the  packing  material  surface.  In  such  cases,  remove  guard 
column and set analysis column with reverse direction with its flow direction, cleaning the column with 50mL 
or  more  of  acetonitrile  or  methanol  may  be  effective  for  recovery  of  column  efficiency.  Guard  column   
should be cleaned alone with the same way. The flow rate should be applied lower than 0.3mL/min. 