background image

 

6H15-019 Rev.3 
Issued  : Jul. 2015 
Revised : Apr. 2019 

 

 

 

 

 

 

Calibration material 

JIS SCM440 flat surface 

1. Calibration material 

MODEL FK-602F Transducers are calibrated for JIS SCM440 flat 
surface (more than 54 mm dia.). 
If the measured target is other than JIS SCM440 flat surface, it will 
present a different characteristics. In such a case, calibration by the 
connected equipment (e.g. monitor) side should be required for 
system operation. 

2.  Shield wire connection 

Connect shield wire of signal cable (3-wire shielded cable between 
driver and monitor) to driver’s “COM” terminal (Spring lock terminal: 
“Shield” terminal) and monitor’s “COM” terminal.   
If this is not adhered to, noise may be caused. 

3. Connector isolation, etc. 

The connector connecting the sensor cable and the extension cable 
shall be insulated with the attached insulation sleeve (transparent 
shrink tube) or fluoro resin insulation tape. 
The vinyl-insulating tape shall not be used. 

4.  Megger test of signal cable 

If megger test is made on the signal cable (3-wire shielded cable), be 
sure to discharge the charged electric load before connecting the 
cable to driver. 
If this caution is not adhered the driver could be dameged. 

5. Sensor installation 

Not available for rain water at out door use. 
It may cause the sensitivity change and insulation down. 

6.  Calibrated as a system 

The sensor, extension cable and driver, which are calibrated as a 
system, shall be connected with each serial No. as specified in the 
inspection test report. If this is not adhered the output characteristics 
may be out of specification. 

7. Linearity  

The linearity margin provides for examination result in our factory. This 
regulated value is not applied to the examination result in the site. 

8.  The instructions manual contains important information such as 

conditions necessary for safe handling of the system. 
Such information and conditions are important and indispensable for 
ensuring safety. Therefore, be sure to read the instructions manual   
thoroughly before handling the system. 

9.  Cable length 5.0m sensor is designed for 5m system only. 

Can not use for 9m system. 

 
 

Measuring range 

0.5 mm to 6.5 mm from sensor tip 

Sensitivity*1 

2.5 V/mm 

Sensitivity error*1 

Within ±4 % 

Linearity*1 Within 

±70 

m of 2.5 V/mm straight line : 

(if calibrated as a system)

Within ±90 

m of 2.5 V/mm straight line :   

(including interchangeability errors)

Linear range : 6 mm 

Frequency responce*1 

DC to 10 kHz or more (-3 dB) 

Max. output voltage*1 

Approx. -23 VDC 

Sensor abnormal output 

voltage*1 

Approx. -0.6 VDC (Sensor OPEN/Sensor SHORT)

Output impedance*1 

50 

 Current 5 mA (max.) 

Current consumption 

(10 k

 load)

 

Max. -15 mA

 

Output noise*1 

Approx. 20 mVpk-pk + power supply noise 

Sensor tip diameter 

Approx. 18 mm dia. 

Cable diameter 

Approx. 3.6 mm dia. 

Connector diameter 

Approx. 7.1 mm dia. 

System cable length

 

5 m or 9 m 

Sensor cable color 

Blue 

Extension cable color 

Black 

Operating temperature 

range

 

Sensor   

-40 to +125 °C

Extension cable 

-40 to +125 °C 

Driver 

 

 

-40 to +80 °C 

Temperature characteristic 
(Temperature drift) 

Sensor 

: Less than ±3 % of F.S.

Extension cable  : Less than ±3 % of F.S. 

  Condition : Gap=5 mm, Target : JIS SCM440 

 

0 to 80 °C (at 20 °C standard) 

Driver 

: Less than ±3 % of F.S. 

Loop 

: Less than ±4 % of F.S. 

  Condition : Gap=5 mm, Target : JIS SCM440 

 

0 to 60 °C (at 20 °C standard) 

Operating humidity range  30 to 95 % RH (non-condensing, non-submerged)

(Sensor body : 100 % RH) 

Power supply 

-24 VDC ±10 % 

Dielectric strength of driver  Between each terminal and mounting plate : 

1 mA or less at 500 VAC for one minute

Insulation resistance of 

driver

 

Between each terminal and mounting plate : 

100 M

 or more at 500 VDC

Applicable wire 

Screw type terminal block (M4)    : 0.75 to 2mm

2

 

Size 

Spring lock terminal            : 0.2 to 1.5mm

2

Driver mass 

Approx. 200 g 

Oil leakage reduction 
performance*2 

Even if oil is poured in to the cable with the 
pressure of 0.05MPa. Oil does not leak from 
opposite the cable for 7 days. 
(cable length : 4m, at 25 °C) 

*1  The above specification apply at 25 °C with -24 VDC power supply and

load resistance 10 k

 and JIS SCM440 target (thickness

5 mm). 

*2 Oil leakage reduction performance is applied to the extension cable part.

 

 

 

 

 

 

 

Specifications 

Notice 

Configuration 

FK-602F Transducer 

Oil leakage reduction extension cable option 

Page 1 of 2

   

 

FK Series 
Transducer 

Specifications

 

Connector 

Sensor 

Extension cable*2 
(Oil leakage reduction)

Driver 

System cable length 
 

FK-602F1 : 5m 

 

FK-602F2 : 9m 