background image

31

7.  Menu Item Selection

7.3 Menu Item Selection Procedure

This instruction manual identifies each menu item by a number. For example, the menu items of “Sta-
bility Detection Band ” of  “11. Environment” are 

 through 

Find the function to be programmed in the Menu Map, referring to the item number in square, 

To reach the item, operate the keys on the balance.

(1) Press the 

[CAL]

 key to cycle through the items within a menu level.

 (

In the Menu Map)

(2) Press the 

[O/T]

 key to choose the current item or move to the next menu level. 

(

In the Menu Map)

(3) Press the 

[POWER]

 key to move back one menu level.

(4) Press and hold the 

[POWER]

 key to return to the gram-display.

Example

: Select “Stability Detection Band” “4 counts”.

The menu item number is 

 on the Menu Map.

1

Press the 

[CAL]

 key repeatedly from the gram-

display until “SEL:EAUS” and some symbols are
displayed and “E” blinks.

2

Press the 

[O/T]

 key. “E” is selected and display

changes to “

E-AbtP8” and “

” blinks.

3

Press the 

[CAL]

 key twice to make “b” blink.

4

Press the 

[O/T]

 key. “b” is selected and “Eb-1”

is displayed. Stability mark is lit if Eb-1 is currently
set.

5

Press the 

[CAL]

 key twice. “Eb-4” is displayed.

   Note

The 

 symbol is displayed during Menu Item selection.

27

33

No.

29

   Note

Before entering the menu, set the balance to the gram-display using the

[UNIT]

 key. It is also possible to enter the menu from other weighing units

involving no further setting with the 

[UNIT]

 key.

Summary of Contents for UW2200HV

Page 1: ...29 21E Oct 2008 For Basic Operation Electronic Balance Instruction Manual UW series UX series Read the instruction manual thoroughly before you use the product Keep this instruction manual for future reference ...

Page 2: ......

Page 3: ... whole or part of the content without permission of the company are strictly prohibited Microsoft Windows Windows Vista and Excel are registered trademarks of Microsoft Corporation of the U S A in the United States and other countries All other company names and product names that appear in this manual are trademarks or registered trademarks of the companies concerned Note that and indications are...

Page 4: ...he weigh ing units These sections include information to make using the balance more convenient Indicates the menu item to be selected The number in the is the number of the menu item on the Menu Map See 7 2 Menu Map Notes on the use of verified balances as legal measuring instruments Important notes about the use of verified balances as legal measuring instruments in the EU are highlighted with t...

Page 5: ...its can lead to accidents as a result of weighing errors Check that the weighing units are correct before starting weighing Do not use the balance outdoors or anywhere where it will be exposed to water You could sustain an electric shock or the product could operate abnormally The levels of danger and damage that will arise if the balance is used incorrectly are classified and indicated as shown b...

Page 6: ...ly Therefore begin operation from 4 4 Turning ON the Power page 18 again Treat the balance with care and respect The balance is a precision instrument Subjecting it to impacts could cause it to fail When moving the balance remove pan and pan supporter Grasp it firmly with both hands to carry it If the balance has to be stored for a long time store it in the packaging box in which it was delivered ...

Page 7: ...ective 89 336 EEC amended by 92 31 EEC 93 68 EEC EN55022 1994 A1 1995 A2 1997 Class B EN55024 1998 EN61000 3 2 1995 A1 1998 A2 1998 EN61000 3 3 1995 Low Voltage directive 73 23 EEC amended by 93 68 EEC EN60950 1992 A1 1993 A2 1993 Weighing Instruments Department Analytical Measuring Instruments Division SHIMADZU CORPORATION 1 Nishinokyo Kuwabaracho Nakagyo ku Kyoto shi 604 8511 JAPAN ...

Page 8: ... VI ...

Page 9: ... VII ...

Page 10: ... VIII ...

Page 11: ...mpre hensive laboratory data management system Shimadzu CLASS Agent Ask your Shimadzu representative about it Shimadzu WindowsDirect When Shimadzu balances are integrated with laboratory software by means of our WindowsDirect function no communication software is required or used The Shimadzu balance functions as a primary device in the system just as a keyboard mouse or other data entry hardware ...

Page 12: ...equip ment is for industrial professional use only Contact Shimadzu service representative when the equipment has reached the end of its life They will advise you regarding the equipment take back With your co operation we are aiming to reduce contamination from waste electronic and electrical equipment and preserve natural resource through re use and recycling Do not hesitate to ask Shimadzu serv...

Page 13: ... the Unit Display 23 6 WindowsDirect Function 24 6 1 Introduction Experience it 24 6 2 Set Up WindowsDirect 24 6 2 1 Setting Up the Balance 24 6 2 2 Cable Connection 25 6 2 3 Setting Up the Computer 25 6 2 4 Start and Checking Operation 27 6 3 Troubleshooting the WindowsDirect Communication Function 28 7 Menu Item Selection 30 7 1 What is the Menu 30 7 2 Menu Map 30 7 3 Menu Item Selection Procedu...

Page 14: ...ng 44 10 3 1 Selecting the Calibration Type 44 10 3 2 PSC Fully automatic Calibration UW series only 44 10 3 3 Clock CAL Fully automatic Calibration UW series only 45 10 3 4 PCAL Calibration of the Built in Weight UW series only 46 10 3 5 PCAL Password Setting UW series only 47 10 4 For GLP GMP ISO Conformance 48 10 4 1 Calibration Report Setting 48 10 4 2 Balance ID Setting 48 11 Environment 49 1...

Page 15: ...l Timer 71 14 5 Add on Mode 72 14 6 Animal Weighing 74 14 7 Formulation Mode 77 15 Connecting Peripheral Instruments 79 15 1 Electronic Printer 79 15 2 Personal Computer RS 232C 80 15 2 1 Connecting the Cable 80 15 2 2 Data Format 81 15 2 3 Using Command Codes 83 15 2 4 Multi Connection Mode 88 15 3 Communication Setting 91 15 3 1 Overview 91 15 3 2 Handshaking 91 15 3 3 Format 92 15 3 4 Communica...

Page 16: ...LCD Liquid Crystal Display Check 97 Appendices 98 A 1 Menu Map 98 A 2 Standard Accessories and Maintenance Parts List 103 A 3 Optional Accessories List 104 A 4 Specifications of Connectors 105 A 5 Table of Unit Conversion Constants 106 A 6 Performance Checks 107 A 7 Below Weigh Hook Dimensions 108 A 8 Index 109 ...

Page 17: ...tor driven built in calibration weight that can automatically calibrate sensitivity without the use of external weights Read this manual carefully before using this instrument and keep it with the balance for future refer ence This manual refers to the different types of UW and UX series UW UX series balances as follows H type UW UX H S type UW UX S Where represents the figure indicating the capac...

Page 18: ... necessary operation and error message Operation keys Used to tare execute calibration and functions program functions input numerical values Level screws Adjust to level the balance Pan Supports the object to be weighed Label Shows model name Max Min e and d are indications required by legal metrology which do not restrict the weighing range in general use Main body Display panel Shows weighed re...

Page 19: ...al metrology which do not restrict the weighing range in general use Main body Display panel Shows weighed result information on programming function indicates current setting working function necessary operation and error message Operation keys Used to tare execute calibration and functions program functions input numerical values Level screws Adjust to level the balance Below weigh hook cap RS 2...

Page 20: ...nstrument in the EU 7 Not applicable to a verified balance as a legal measuring instrument in the EU Key During Weighing Press Once and Release Press and Hold for About 3 Seconds POWER Switches between the operation and standby modes Exits the application function and returns to the mass display CAL Enters span calibration or menu item selection 1 Displays the last menu item that was set Last menu...

Page 21: ...es that the menu lock is on Illuminates during menu item selection Asterisk Indicates that the displayed numeric value is not a mass value Communication symbol Illuminates during communication to external equipment through the RS 232C or DATA I O connector In menu selection indicates setting related to communication Battery symbol When the balance is operated with the optional battery pack this sy...

Page 22: ...C Clock CAL GLP GMP ISO conformance Analog display display PCS User unit Animal weighing Specific gravity measurement S W Checkweighing UX Series Model UX220H UX420H UX620H UX2200H UX4200H UX6200H UX420S UX820S UX4200S UX8200S Capacity 220g 420g 620g 2200g 4200g 6200g 420g 820g 4200g 8200g Minimum display 0 001g 0 001g 0 001g 0 01g 0 01g 0 01g 0 01g 0 01g 0 1g 0 1g Calibration range with external ...

Page 23: ...racy class II Capacity N A 1100ct 2100ct 3100ct N A 11000ct 21000ct 31000ct Verification scale interval e N A 0 1ct 0 1ct 0 1ct N A 1ct 1ct 1ct Number of verification scale interval N A 11000 21000 31000 N A 11000 21000 31000 Scale interval d N A 0 01ct 0 01ct 0 01ct N A 0 1ct 0 1ct 0 1ct Range of use N A N A 0 2ct 1100ct 0 2ct 2100ct 0 2ct 3100ct N A N A 5ct 11000ct 5ct 21000ct 5ct 31000ct Tare r...

Page 24: ...ched AC adapter is used properly If this is not possible an optional battery pack is available as a special accessory Verify that the supply power voltage conforms to that indicated on the AC adapter 2 Installation site Air flow from air conditioner open window or venti lator Vibration Direct sunlight Continued Caution Avoid sites where the balance will be exposed to the following ...

Page 25: ...agnetic waves or magnetic fields Large capacity balances should be installed on a sturdy floor and table that can support the total load of the balance AND object to be weighed Note Using a verified balance as a legal measuring instrument in the EU The balance must be used within the temperature range indicated on the verification label ...

Page 26: ...play 0 01g c Small pan model Minimum display 0 001g Model UW UX is UW or UX Additional suffix may appear after H or S on your balance UW UX2200H UW UX4200H UW UX6200H UW UX4200S UW UX8200S UW UX420S UW UX820S UW UX220H UW UX420H UW UX620H Balance main body 1 1 1 Pan supporter cap 4 4 4 Pan 1 1 1 AC adapter 1 1 1 In use protective cover 1 1 1 Windbreak set Main 0 0 1 Lid 0 0 1 Fixing knob 0 0 2 Rub...

Page 27: ...er from this figure Instruction manual Explanatory operation sheet Pan Pan supporter caps Balance main body In use protective cover Stainless screws Instruction manual Explanatory operation sheet Pan Pan supporter caps Balance main body with rubber caps In use protective cover AC adapter The shape of the adapter supplied with the balance may differ from this figure ...

Page 28: ...reak Main Windbreak Lid Stainless screws Fixing knobs for windbreak AC adapter Instruction manual Explanatory operation sheet Pan Pan supporter caps Balance main body with rubber caps In use protective cover The shape of the adapter supplied with the balance may differ from this figure ...

Page 29: ...on the bottom of the balance turn the two transportation screws counterclockwise until they tighten again UW only Caution Do not operate step 2 with the balance placed on its side Place the balance on a smooth surface Transportation screws only for UW series Below weigh hook cap Explanation label only for UW series Caution When moving the balance again turn the two transportation screws clockwise ...

Page 30: ...ont corner of the balance 2 so that both front level screw feet 3 are touching the table surface 3 Bubble moves to the highest position There fore adjust level screws 3 so that the balance main body is lowered in the direction of the bubble Case 1 Right front of the balance is too high Turn right front level screw anti clockwise so that the bubble moves towards center Case 2 Front of the balance i...

Page 31: ...l minimum display 0 01g Insert the four pan supporter caps into the holes in the top of the balance Place the pan gently on pan supporter caps Positioners of the pan must fit pan supporter The rubber caps on top of the main body may be replaced with the stainless screws so that it will be more secure when exposed to organic solvent c Small pan model minimum display 0 001g windbreak standard 1 Pull...

Page 32: ...sten it with two fixing knobs 3 Insert the four pan supporter caps into the holes in the top of the balance Place the pan on them Positioners on the pan must fit pan supporter caps 4 Place windbreak lid on top of windbreak main fitting the hinge parts Curved front Pan Pan supporter caps Hinge parts ...

Page 33: ... adhesive parts gently Note Using a verified balance as a legal measuring instrument in the EU Legal regulations require a verified balance be sealed This control seal is a self destructive adhesive label This seal is irreparably damaged invali dating the verification if you attempt to remove it The balance must then be re verified before it is used for legal measurements ...

Page 34: ...ighting oFF CHE 5 and CHE 4 are not displayed for the UX series 3 Press POWER key The whole display illumi nates and then the display changes to indicate the gram display The backlight is illuminated Note A power cable may be necessary to connect the AC adapter to the power source depending on the type of the AC adapter AC Adapter Balance Power Source Power cable Note When using the optional batte...

Page 35: ...lay and that the pan is empty 2 Press the CAL key once i CAL is displayed 3 Press the O T key After i CAL3 i CAL1 Set CALEnd are displayed indicating the com pletion of span calibration the gram display will appear This is the standard calibration type Refer to 10 3 1 for use of external weights Note Using a verified balance as a legal measuring instrument in the EU Span calibration must be perfor...

Page 36: ...ded is displayed and blinks 4 Load the indicated calibration weight and press the O T key 5 When the zero display blinks unload the weight from the pan and press the O T key Set is dis played briefly to indicate completion of span cali bration Then the gram display will return Note Span calibration should be performed again when the location of the balance is changed when the room temperature chan...

Page 37: ...or Basic Operation Note Span calibration should be performed again when the location of the balance is changed when the room temperature changes considerably periodically according to the quality control plan of the user ...

Page 38: ...ate 2 Press the O T key to zero the display This operation is called taring 3 Place the object to be weighed on the pan 4 Read the displayed value after the stability mark is displayed Note Using a verified balance as a legal measuring instrument in the EU Indicates that the balance is set exactly to Zero with the zero setting function 0 20e e verification scale interval ...

Page 39: ... verification label When PSC refer to 10 3 2 fully automatic span calibration is not acti vated operator must carry out span calibration with the built in weight refer to 4 5 upon blinking of the Weight Symbol Overload Weighing capacity has been exceeded Negative Overload The load on the balance is too light The pan is not adjusted properly For D type balances oL will appear if the load is below t...

Page 40: ... software is used WindowsDirect function should be turned off Set up the opti mal communication parameters for the software according to 15 3 Communication setting 6 2 1 Setting Up the Balance 1 Press the MENU key twice from the gram dis play Setwin appears 2 Press the O T key Verify the stability mark is illuminated with Setwin display All the communi cation settings for WindowsDirect have been m...

Page 41: ... and start Windows 2 Click Start choose Settings and Control Panel 3 Select Accessibility Options 4 Verify that there are no check marks for any items on all five tabs including General RS 232C Connector When using WindowsDirect use a Null modem cable of one of the below wirings 1 2 D sub9 1 2 3 4 5 6 7 8 9 D sub25 1 2 3 4 5 6 7 8 20 22 D sub9 2 3 4 5 6 7 8 D sub25 2 3 4 5 6 7 20 is not necessaril...

Page 42: ...nistrative options appears in the General tab Put check marks at both the items of Administrative options to maintain the settings even after restarting Windows 6 Open Settings 7 Select the serial port corresponding to the RS 232C port of your personal computer Serial port any one of COM1 to 4 Usually COM1 8 Select a Baud rate of 300 9 Click OK 10 Click Apply and wait 11 Click OK ...

Page 43: ...e numeric value displayed on the bal ance appears at the cursor position on the screen of computer The effect is the same as typing the value from the computer keyboard and pressing the ENTER key Characteristics indicating the unit of measure are not sent to the computer 5 Test combination with Auto Print function if you wish to use it Refer to 13 3 6 End the operation using the standard close or ...

Page 44: ...erly Try uninstalling the driver and installing it again Some notebook PCs feature a setting for disabling RS 232C ports as a power saving measure Before trying to use the WindowsDirect communication function make the setting that enables the use of RS 232C ports Communications with other applications and PCs via a LAN may interfere with the serial key device settings Try using WindowsDirect commu...

Page 45: ...ata in another application e g Notepad Q6 The operation is sometimes abnormal A6 Depending on the processing capability of the PC malfunctions may occur if the communications speed is high Set 300 bps for the communication speed Malfunctions may also occur if the interval for data transmission from the balance is too short Ensure that one batch of data is displayed on the screen before the next ba...

Page 46: ...er Menu Item selection is used to program these functions 7 2 Menu Map The menu of the UW UX balance consists of seven groups and four levels The Menu Map shows the structure clearly with menu item numbers to help access the desired function Refer to the Menu Map on the operation explanatory sheet or in appendix when programming the functions in Chapter 8 through 15 ...

Page 47: ... and hold the POWER key to return to the gram display Example Select Stability Detection Band 4 counts The menu item number is on the Menu Map 1 Press the CAL key repeatedly from the gram display until SEL EAUS and some symbols are displayed and E blinks 2 Press the O T key E is selected and display changes to E AbtP8 and blinks 3 Press the CAL key twice to make b blink 4 Press the O T key b is se...

Page 48: ...he power is disconnected Major Menu Description Important Note on Menu Item Selection Even the desired menu item is reached and displayed it is not yet set unless Stability mark is illuminated with it Do not fail to press O T key to put Stability mark before returning to the mass display Menu Group Symbol that blinks at beginning of menu Menu Items Included 1 Calibration 2 Graphic display Analog d...

Page 49: ...sfully saved Err is displayed when the balance failed to save the value 4 Press the POWER key to stop numeric entry Abort is displayed briefly and the display returns to the menu one level up Use the optional AKB 301 Application Keyboard to easily set numerical values and decimals Notes Setting a Decimal Point A decimal point is only used when setting units for solid density weighing liquid densit...

Page 50: ... to reset the menu 1 In the gram display press the CAL key repeatedly until the S of SEL EAUS blinks 2 Press the O T key The Menu Group 6 is selected 3 Press the CAL key repeatedly until the r in S dtSCr is blinking 4 Press the O T key to display rESEt without the dot 5 Press the O T key again rESEt is displayed to indicate menu reset completion 6 Press the POWER key several times or hold it for a...

Page 51: ... When oFF is displayed press and hold down the CAL key for about three seconds 3 rELEASE is briefly displayed to indicate that the menu lock has been turned off How to lock the menu Notes is illuminated during oFF display or STAND BY while Menu Lock is activated LoCKEd is displayed upon an attempt of access to the menu including releasing the currently set function while Menu Lock is activated All...

Page 52: ...ted from three styles The setting made here is not reflected on the display of the balance To output in the YYYY MM DD order select menu item y m d To output in the DD MM YYYY order select menu item d m y To output in the MM DD YYYY order select menu item m d y 63 Example Notes The built in clock corrects for the leap year automatically The moment the O T key is pressed to finish setting seconds a...

Page 53: ...s 13 23 8 4 Setting Display During Stand by Determine what is to be displayed during stand by To display the time during stand by select menu item To display the date during stand by select menu item To display neither during stand by select menu item 64 Example Note The moment the O T key is pressed seconds are set to 00 Notes Seconds display function Press the UNIT key to switch on off of the di...

Page 54: ...s that the load on the pan is close to the weighing capacity 2 To display no bar graph select menu item 9 2 Changing the Minimum Display Digit 10d 1d Not applicable to a verified balance as a legal measuring instrument in the EU It is possible to decrease the resolution of the minimum balance display by one decimal place if neces sary 1 Press and hold the UNIT key for approxi mately three seconds ...

Page 55: ...e user Terms used in this manual Span Calibration Adjustment of the balance to specifications using two weight values zero and an appropriate value for the balance capacity Calibration Check Comparing the current calibration weight reading to the calibration weight reading after the last span calibration Calibration Pertains to both span calibration and calibration check Caution Never plug off the...

Page 56: ...splayed If i CAL is not displayed return to mass display and select menu item 3 Press the O T key After i CAL3 i CAL2 i CAL1 Set CALEnd the mass display will appear indicating the com pletion of span calibration Notes Setting before shipment is as the following UW series Span calibration using the built in weight UX series Span calibration using external weights The type of calibration can be chan...

Page 57: ...t calibration weight reading and the cal ibration weight reading at the last span calibration 4 To perform span calibration change the d value to zero by pressing the CAL key Otherwise Press the O T key to avoid chang ing the d value to zero Pressing the POWER key interrupts calibration and does not change this value to zero CALEnd is displayed indicating the completion of the calibration check 2 ...

Page 58: ...calibration weight and press the O T key 5 Shortly zero display blinks Unload the weight from the pan and press the O T key SEt is displayed briefly to indicate completion of span calibration Example 3 Changing the Calibration Weight to be Used Pressing the CAL key allows changes to the weight value Modify the value using the UNIT key and PRINT key then press the O T key Refer to 7 4 for how to ma...

Page 59: ...display xxx indicates a numeric value 6 To perform span calibration change the d value to zero by pressing the CAL key Otherwise press the O T key to avoid changing the d value to zero Pressing the POWER key interrupts calibration and does not change this value to zero CALEnd is displayed indicating the completion of the calibration check 4 Example Changing the Calibration Weight to be Used Pressi...

Page 60: ...10 3 2 PSC Fully automatic Calibration UW series only With the PSC function span calibration is performed automatically using the built in calibration weight when the balance detects a temperature change that would affect weighing accuracy 1 To turn ON the PSC function Select menu item 2 To turn OFF the PSC function Select menu item 1 2 3 4 5 6 Notes Blinking calibration symbol indicates an automa...

Page 61: ...the O T key The set time will be stored 4 Proceed to next time setting by the MENU key or return by the POWER key Skipping Clock CAL If Clock CAL starts while the balance is in use press the POWER key to abort that cycle Turning Off Clock CAL Function When all three Clock CAL times are set to 00 00 the function is off Notes To execute Clock CAL all of the following conditions must be satisfied at ...

Page 62: ...PAS 0000 is displayed 3 Enter the PCAL password using the UNIT and PRINT keys then press the O T key The default password is 9999 set at shipment or upon menu reset After PCAL 3 is displayed the value of the stan dard weight to be loaded blinks 4 Load the standard weight displayed and press the O T key Soon zero is displayed and blinks 5 Unload the weight and press the O T key The display proceeds...

Page 63: ...event an unauthorized person from incorrectly calibrating the built in calibration weight Select menu item The numerical setting display appears Enter a 4 digit number from 0000 to 9999 Refer to 7 4 for numerical input Notes SEt is displayed during the process Leave the balance in a stable state until the mass display appears as in step 5 In PCAL the value of the weight to be loaded cannot be chan...

Page 64: ...ptional accessory is required to print the report To create calibration report Select menu item To turn off calibration report function Select menu item 10 4 2 Balance ID Setting Individual balances can be identified by the serial number on the main body of the balance The user can add a four digit ID number to the calibration report Select menu item Set a 4 digit number from 0000 to 9999 Note Dat...

Page 65: ...unless special circumstances exist Pouring mode Select menu item Filling dosing This mode is particularly suitable for dosing or filling purposes Note that this mode is very sensitive to wind and vibration Environmental setting in Pouring mode Pouring mode allows further adjustments to the surrounding environmental conditions Choose the optimal setting for your installation site observing the resp...

Page 66: ...e following purposes 1 When stability is detected during measurement Stability mark is illuminated as an auxiliary indi cation of measurement stability 2 Operations such as data output and auto zeroing of the below listed functions are triggered by stability detection Some data output functions operated with communication commands 15 2 3 are also linked to stability detection Functions that employ...

Page 67: ... be regarded to be stable when the display has remained within two display counts Selecting a larger stability detection band generally eases the criteria and makes stability mark illumi nate more readily However it will be more likely for the display to further fluctuate after illumination of stability mark Try a larger stability detection band when data outputs triggered by stability detection i...

Page 68: ...stability mark select menu item Select menu item to return it to the standard setting 11 4 Tracking Tracking is the function that will maintain the current displayed value as long as possible To turn ON this function Select menu item To turn OFF this function Select menu item Zero Tracking Function When the display is zero tracking functions as zero tracking to keep the zero display as long as pos...

Page 69: ... Error Display Note Using a verified balance as a legal measuring instrument in the EU Only g gramme gram ct carat weighing units percentage conversion and piece counting are available for use Carat is not available for UW820SV and UW8200SV models Example When menu item carat has been selected 58 54 62 Notes For the unit names refer to the menu map In the unit setting menu the stability symbol is ...

Page 70: ...s to the 100 value This value must be equivalent to 100 counts or more in the g unit 3 When the stability mark illuminates press the CAL key SEt is displayed briefly and the reference sam ple weight is displayed as 100 The weights of subsequent samples are dis played as a percentage of the reference sample weight 56 Note The numbers of digits displayed in the unit and rounding off of the mini mum ...

Page 71: ...target mode and full scale mode Only one of them can be used at a time Refer to 9 1 for the full scale mode and the no bar graph display Upon selection of either the checkweighing mode or the target mode the numerical value for threshold or target is requested Numeric values are memorized independently for each graphic display mode When one of the checkweighing modes is selected a signal correspon...

Page 72: ...o includes a checkweighing function 1 Select menu map item 2 Set the upper threshold value which corresponds to the upper triangle mark with menu item 3 Set the lower threshold value which corresponds to the lower triangle mark with menu item Displays in use 15 16 17 Note The decimal point is invisible in each value setting display Setting the value cannot be made in gram unit Determine the number...

Page 73: ...orresponds to the center line of the graphic display with menu item 3 Set the limit value which corresponds to the distance between the center line and upper or lower triangle marks with menu item Note The decimal point is invisible in each value setting display Setting the value cannot be made in gram unit Determine the number based on the minimum display of the balance For example if the balance...

Page 74: ... Every time the CAL key is pressed the dis play sequentially changes as Ld 5pcs Ld 200pcs Ld 5pcs The default setting is Ld 10pcs Press the O T key when the display is equiva lent to the number of loaded pieces Example If 50 pieces are loaded press the O T key when Ld 50pcs is displayed This determines the unit weight or average weight per piece As sample is added or removed the piece count number...

Page 75: ...lue within the Zero Range by unloading the sample or pressing the O T key Print on loading and on zero Select menu item Load the sample when the value displayed is within the Zero Range When stability is detected and the positive displayed value is more than 5 times the Zero Range data is automatically output Unload the sample or press the O T key When the displayed value is within the Zero Range ...

Page 76: ...ity detection Pretaring Value 13 7 function cannot be used with Auto Print functions Auto Print mode setting can be cleared without menu operation by pressing the POWER key for about three seconds Not more than one of these seven functions can be set at a time Setting the second one with the menu will automatically clear the previously set function Auto Print 13 3 Auto Zero 13 4 Peak Hold 14 3 Int...

Page 77: ...of stability detection Pretaring Value 13 7 function cannot be used with Auto Zero function Auto Zero setting can be cleared without menu operation by pressing the POWER key for about three seconds Not more than one of these seven functions can be set at a time Setting the second one with the menu will automatically clear the previously set function Auto Print 13 3 Auto Zero 13 4 Peak Hold 14 3 In...

Page 78: ...t sample If it does not return below 0 10g the previous sample is not judged to be unloaded Therefore even when the displayed value reaches 0 50g again by next loading the data will not be automatically output Select Zero Range value according to the influences of environment causing difficulty in zero return the weight of the samples etc so that Auto Print functions properly Set the Zero Range va...

Page 79: ...cted Waiting for stability detection Select menu item Notes While the balance is waiting for stabilization is displayed When the O T key has been pressed is displayed At this occasion press the POWER key if you desire to disable this function and abort taring When the PRINT key has been pressed Once the communication symbol and the Stand by symbol are lit then stability detection is waited The dat...

Page 80: ...f the sample 1 Select menu item 2 Set the Pretare value Refer to 7 4 for numeri cal value input Notes If the weight of the tare container varies accurate measurement with Pretaring Value function cannot made Pretaring Value function cannot be used with Peak Hold 14 3 Add on Mode 14 5 Animal Weighing Mode 14 6 Formulation Mode 14 7 or Auto Print function 13 3 36 Pretare Value Cancel the Pretare val...

Page 81: ...pecific gravity measurement refers to the measurement of the sample solid weight in the air and in a liquid of known specific gravity or density and the calculation of the sample specific gravity or density The Td symbol is used to represent the solid specific gravity in this balance The data output unit is DS 1 Select menu item in advance 2 Remove the below weigh hook cap from the bottom of the b...

Page 82: ...p to four decimal places are displayed for density or specific gravity Depending on the measurement conditions not all digits are stable even the balance is working normally When it is not possible to stabilize the display in all 4 decimal places you may change the minimum display by using the UNIT key How to change the minimum display in density specific gravity display When the result is display...

Page 83: ...ok 3 Hook the hanging pan and then immerse the hanging pan in a tank containing the sample liq uid 4 From mass display press the UNIT key sev eral times until d is displayed 5 Press the O T key 6 Load the reference weight on the pan of the balance Note Use of the optional SMK 101 or SMK 102 Specific Gravity Measure ment Kit refer to A 3 is recommended for efficient measurements When using the SMK ...

Page 84: ...igits are stable even the balance is working normally When it is not possible to stabilize the display in all 4 decimal places you may change the minimum display by using the UNIT key How to change the minimum display in density specific gravity display When the result is displayed press and keep holding down the UNIT key for about three seconds This changes the minimum display to 10d d original m...

Page 85: ... advance 1 In the peak detection standby state with the P symbol and the Stand by symbols illuminated press the O T key to tare the display 2 Press the PRINT key The Stand by symbol disappears and peak value detection starts 3 Upon stability is detected the operation of peak value detection is ceased The detected peak value is displayed with the P symbol and Asterisk simultaneously illuminated The...

Page 86: ...g menu item Pouring mode Depending on the weighting conditions and the sample type this setting may not always be suitable Pretaring Value 13 7 function cannot be used with Peak Hold function See 13 5 for details of Zero Range See 11 3 for details of stability detection Not more than one of these seven functions can be set at a time Setting the second one with the menu will automatically clear the...

Page 87: ...Use the O T key to erase the tare or zero the balance at any time Pressing the POWER key in the interval timer standby state brings the power supply standby state Interval Timer setting can be cleared without menu operation by press ing the POWER key for about three seconds This does not clear the set time interval value Using the interval timer function to record data over a long period may cause...

Page 88: ...ey in the Add on standby state The Add on and the Standby symbols are lit Zeroing occurs 2 Press the PRINT key The Stand by symbol disappears and Add on measurement starts 3 Load the first sample When stability is detected with a displayed value five times the Zero Range or more the displayed value is automatically output and zeroing occurs Even if the displayed value is less than five times the Z...

Page 89: ...e See 11 3 for details of stability detection Weighing unit 12 1 and minimum display 9 2 cannot be changed when Add on Mode is on When Add on Mode is ON fully automatic span calibration by PSC 10 3 2 or Clock CAL 10 3 3 is not performed Weight symbol keeps blinking when span calibration is necessary Span calibration with the built in calibration weight or external weights can be performed between ...

Page 90: ...ponse and stability of measurement Operational Conditions Cond 2 and Cond 3 are not suitable for weighing animals lighter than the respectively specified in the above When lighter animals are weighed with these settings the zero return after removing the animal may be very slow Switching operational conditions Every time the UNIT key is pressed at the zero display in Animal Weighing mode the opera...

Page 91: ...2 Load the sample animal with a mass more than 50 times the Zero Range 3 When the measurement reaches stability crite ria of Animal Weighing the weighed value is auto matically output 4 Unload the animal or press the O T key 5 When the mass display becomes less than 10 times the Zero Range and stability is detected automatic zeroing occurs The mass of any residue remaining on the bal ance excremen...

Page 92: ...tional condition of Animal Weighing is also locked When Menu reset 7 5 2 is performed operational condition setting of Animal Weighing is also returned to the default Cond 1 Pretaring Value 13 7 function cannot be used with Animal Weighing mode Small animal bucket set option see A 3 is recommended for weighing small animals such as mice When the POWER key is pressed in the Animal Weighing standby ...

Page 93: ...re illuminated in the display Place a container if used and press O T key to tare the display Note that taring with O T key is not accepted after PRINT key is pressed in the following step 2 until POWER key is pressed in the step 5 2 Press PRINT key When external device is connected FORMULATION MODE will be output 3 Load the first component and press PRINT key Upon stability is detected at a value...

Page 94: ...ng 13 7 Weighing unit 12 1 and minimum display 9 2 cannot be changed when Formulation Mode is on When Formulation mode is ON fully automatic span calibration by PSC 10 3 2 or Clock CAL 10 3 3 is not performed Weight symbol keeps blinking when span calibration is necessary Span calibration with the built in calibration weight or external weights can be performed between sets of measurements Formula...

Page 95: ...1 For the balance select the default communica tion settings 2 Disconnect the power to the balance and the electronic printer 3 Connect the DATA I O connector of the balance to the electronic printer using the attached cable of the printer 4 First turn ON the power to the balance then turn ON the power to the electronic printer Menu item number Setting Handshaking Timer Format EB type Baud Rate 12...

Page 96: ...ub9 pin Null modem Notes The cable must have the correct connections as shown in the diagram below Cables with the connections shown below and the special accessory RS 232 cable are not guaranteed to operate properly for all types of computers and devices Computer side Balance side RXD 2 2 TXD TXD 3 3 RXD DTR 4 6 DSR SG 5 7 SG DSR 6 20 DTR RTS 7 5 CTS CTS 8 4 RTS ...

Page 97: ... used for the unit Position 12 Code for delimiter 2 Information of additional bytes i Output with stability information When outputting data with stability information Refer to 15 2 3 4 i either of the following characters is added in front of Character No 1 in the example Consequently the data becomes 1 byte longer When stable S When unstable U ii Verified balance as a legal measuring instrument ...

Page 98: ...y the data becomes 1 byte longer 3 Data format in case of oL or oL Overload The below is the data format for oL For oL negative overload Position 1 is replaced with minus ASCII code 2DH The following parts appearing in 2 Information of additional bytes also apply to 3 i Output with stability information iv When the delimiter C R L F is selected Data length of this example 12 bytes Position 1 2 3 4...

Page 99: ...Example 5 MENU 4630 CR 4 digit number after This is the same in the following operation from mass display press the CAL key five times then press the O T key once and the press the CAL key six times This returns to the same display as if the CAL key was never pressed Press the O T key once then CAL key three times and the O T key once more This example indicates that menu item Anti vibration mode ...

Page 100: ...ving this command the balance outputs ABCDEFG12345 CR The printer can print this character string 4 Command codes for Format EB type menu item and Format Old EB type menu item i Commands related to output Notes If there is 0 in the four digit number the setting is complete at that point and menu selection is ended The result of this command varies depending on the type of the balance Note Only cap...

Page 101: ...pplication Keyboard PEAK Sets the Peak Hold mode AZERO Sets the Auto Zero mode ON INTERVAL Sets the Interval Timer mode MEMORY Sets the Formulation mode M Immediately operates after setting the Formulation mode ADDON Sets the Add on mode Immediately operates after setting the Add on mode A Sets the Animal Weighing mode ANIMAL Sets the Animal Weighing mode R Releases the Application weighing mode g...

Page 102: ...erence weight set value DENSE Readout of surrounding liquid density set value I TIME Readout of Interval Timer set value CALWT Sets external weights value for span calibration ACALT1 Sets Clock CAL time 1 ACALT2 Sets Clock CAL time 2 ACALT3 Sets Clock CAL time 3 P TARE Sets Pretare value ZRNG Sets Zero Range value UW Sets unit weight USER Sets user unit conversion coefficient VOL Sets volume of re...

Page 103: ...e ADJCLK Adjusts 30 seconds RSTMN Menu reset MENU Sets arbitrary menu Echo back Echo back α Sets to Multi Connection mode α represents a lower case alphabet character Note For Multi Connection mode refer to Section 15 2 4 Multi Connection Mode S One time output at a stable state SI Immediate one time output SIR Continuous output SR Continuous output at a stable state T Taring after stabilized TI I...

Page 104: ... identification name in lower case alphabets Since 10th character of alphabet is j assign j to the balance closest to the personal computer i to next closest balance and so on back to a Setting to Multi Connection Mode 1 Adjust the communication menu settings of all the balances to the specification of the personal computer and then turn off the power to all bal ances 2 Start supplying the power t...

Page 105: ...of date time and cali bration document etc BALANCE No Command PC RETURN DATA PC a a PRINT a PRINT 0 0g b b PRINT b PRINT 0 0g c c PRINT c PRINT 0 0g d d PRINT 0 0g nearest to PC No Data α TARE Same as the operation of pressing the O T key of the balance α α T Same as the operation of pressing the O T key of the balance α α POWER Same as the operation of pressing the POWER key of the balance α α Q ...

Page 106: ...nces in the Application Measurement mode such as Auto Print or Continuous Output or pressing the PRINT key Normal operation cannot be guaranteed Peripheral instruments connected to the DATA I O or IFB 102A connector such as an EP 50 Elec tronic Printer cannot be used The communication formats can only be used with Format EB type Delimiters cannot be used with menu items or setting for WindowsDirec...

Page 107: ...re handshaking performed as the following select menu item After the balance receives X OFF 13H the balance output is retained After the balance receives X ON 11H the balance output is initiated To have hardware handshaking performed as the following select menu item When DTR is OFF the output from the balance is retained When DTR is ON the output from the balance is initiated To have timed hardwa...

Page 108: ...e IS series of Sartorius electronic balances Available commands functions and responses are limited Select menu item 15 3 4 Communication Speed Select the communication speed 300 600 1200 2400 4800 9600 19200 or 38400 bps Number of b xxx shows bps bits second Baud rate and bps are the same value Select one of the menu items to 15 3 5 Parity Bit Length Select the parity and bit length No parity 8 b...

Page 109: ...lue Select menu item 15 4 Decimal Point Symbol in Output Data The decimal point symbols in the outputted data to external devices can be selected from period or comma depending on your preference Note that the decimal point expression on the balance dis play is always with period To select as outputted decimal point select menu item number dECP Pr To select as outputted decimal point select menu i...

Page 110: ...o deliver the balance For UW balances Before placing the balance into the shipping carton remove the pan and pan supporter caps and verify that the transportation screws in the bottom of the balance have been tightened clockwise until they stop Refer to 4 3 For UX balances Remove the pan and pan supporter caps and place the balance in the packing case as it was shipped Caution Failing to tighten t...

Page 111: ...ion Measurement was released Calibration check detects too large error Contact your Shimadzu representative Calibration check detects too large error Contact your Shimadzu representative Menu lock is applied Menu lock is released Menu was reset Contents of new setting and coefficient were stored Reset by power failure Built in weight is moving Wait All numerals blinking Place the displayed calibra...

Page 112: ... Use correct weight The drift is large at span calibration Use correct weight Calibration weight is wrong Use correct weight Failure in the balance Stops when this is displayed Received command code is not cor rect Check delimiter etc Integer of the displayed unit exceeded 7 digits Decrease the load Failure in the balance PCAL password error Check the password Attempted to set improper numeric val...

Page 113: ...ews haven t been loosened UW only Turn them anti clockwise until they stop See 4 3 The pan is displaced Place the pan properly Pan supporter caps are not installed Install pan supporter caps See 4 3 The load on the pan is too large Use balance within its capacity Display does not change when pan is loaded Pan is displaced Place the pan properly Display fluctuates Affected by vibration or air flow ...

Page 114: ...The mass display CAL Execution menu Executes calibration WindowsDirect Setting mark blinking Menu Group 1 CAL key three times from mass display Sets type of calibration to execute blinking Sets I CAL Adjusting sensitivity with the built in weight 1 blinking Sets I TEST Testing sensitivity with the built in weight 2 blinking Sets E CAL Adjusting sensitivity with the external weight 3 blinking Sets ...

Page 115: ...y blinking Stability and Response Timing of Stability Mark Auto 22 Pouring 23 Standard 24 21a 21b 21c Anti vibration 25 Anti wind 26 blinking Stability detection band 1 count 27 2 counts 28 4 counts 29 8 counts 30 16 counts 31 32 counts 32 64 counts 33 blinking Tracking ON 34 OFF 35 and blinking Pretaring Whole lighting at start 36 blinking holds 37 Does not hold 38 blinking Taring Printing at Sta...

Page 116: ...from mass display kg kg kg 54 mg mg mg 55 pcs Number PC 5 5 7 ct Carat CT 58 mom Momme MO 59 d Solid Specific gravity measurement 60 d Liquid density measurement 61 Lb Pound 61a Oz Ounce 61b Ozt Troy Ounce 61c Hong Kong s tael 61d Singapore tael 61e Taiwanese tael 61f Malaysia tael 61g Chinese tael 61h 61i GN Grain 61j m Mesgal 61k b baats 61l t tara 61m Set the Multiplier for User uni t US 62 Dur...

Page 117: ...isplay 67 63a 63b 63c blinking Report and control setting CAL report output On 68 Off 69 Balance ID 70 PCAL Password 71 blinking blinking Menu reset to default Decimal point output 72 Date output style Date setting Period Comma Time setting 72a 72b Continued Menu item number Default settings No Not applicable to a verified balance as a legal measuring instrument in the EU ...

Page 118: ... PR type 79 IPS type 80 blinking Baud Rate Bits sec 300 bps 81 600 bps 82 1200 bps 83 2400 bps 84 4800 bps 85 9600 bps 86 19200 bps 87 38400 bps 88 blinking Parity bit length None 8 bits 89 Odd 7 bits 90 Even 7 bits 91 blinking Stop bit 1 bit 92 2 bits 93 blinking Delimiter CR 94 LF 95 CR LF 96 WindowsDirect down 97 WindowsDirect right 98 Returns to blinking Menu item number Default settings No ...

Page 119: ...r display and key part In use protective cover 5 pieces 321 62395 10 For display and key part Level screws 321 53530 30 Windbreak main 321 62799 01 Standard for models with minimum display of 0 001g only Windbreak lid 321 62798 01 Windbreak fixing knob 1 piece 321 62787 Windbreak Set 321 62795 Set of 321 62799 01 321 62798 01 and 2 pieces of 321 62787 Rubber cap 1 piece 321 62794 For small pan mod...

Page 120: ...y of 0 001g WBC 102 Glass Windbreak 321 62795 For small pan model only with three sliding doors WBC 502 Large Size Windbreak 321 53537 For all models opens upward with hinges FSB 102PK Foot Switch 321 60110 11 The same effect as PRINT key FSB 102TK Foot Switch 321 60110 12 The same effect as O T key Animal Bucket Set 321 62150 For animal weighing for large pan model only Specific Gravity Measureme...

Page 121: ...Handshake transmission 21 NC Blank 22 NC Blank 23 NC Blank 24 NC Blank 25 NC Blank KEY Connector When used for checkweighing output Pin No Name Function Remarks 1 GND Ground of checkweighing 2 OPERA TION Output during weighing Checkweighing output is O C open collec tor Capacity 5 to 24VDC 0 to 20mADC 3 STABIL ITY Output while balance display is stable 10 LO LO signal output 11 GO GO signal output...

Page 122: ...none solid specific gravity ______ ______ 61 U d none liquid density ______ ______ 61a U lb none Lb pound 0 00220462 0 00005 61b U oz none Oz ounce 0 035274 0 0005 61c U ozt none Ozt troy ounce 0 0321507 0 0005 61d U HK tl 1 Hong Kong tael 0 0267165 0 001 61e U HK tl 2 3 4 Hong Kong tael jewel 0 0267173 0 001 61f U SPorE tl 2 Singapore tael 0 0264554 0 001 61g U twn tl 3 Taiwan tael 0 0266667 0 00...

Page 123: ... formulas shown below to calculate the values Rx and Ry 4 Balance repeatability is considered normal when both Rx and Ry are within five counts One count corresponds to one minimum dis play unit of the balance model Notes Performance checks should be conducted in a room where there are no sudden temperature changes Refer to the installation guidelines for the environmental factors that assure opti...

Page 124: ...ally at five different positions on the pan in the order shown Record the readings in each of the five positions 3 If the difference between the readings at the center position and each of the off center posi tions eccentric error is within five counts the balance cornerload performance is considered normal One count corresponds to one minimum dis play unit of the balance model A 7 Below Weigh Hoo...

Page 125: ...ss JSee Calibration weight Calibration report 48 Calibration weight 20 39 42 43 44 45 46 47 Checkweighing 55 56 60 Clock 36 45 Clock CAL Fully automatic Calibration 45 Command code 83 84 89 Communication speed 92 Comparator 56 Computer 25 88 Connector 3 105 Continuous output 59 D Data format 92 Date 36 37 86 Decimal point 33 93 Default setting 34 Delimiter 93 Delivery 10 Density 65 67 Display 5 37...

Page 126: ...rals JSee Optional accessories Personal computer 88 Piece counting 53 58 Pouring mode 49 Power source 8 18 Pretare 64 86 Printing JAlso See Output 63 R Readability JSee Minimum display Remote display 104 Response 49 S S type 1 Sensitivity adjustment JSee Span calibration Small pan model 1 Span calibration 19 20 39 40 42 44 45 86 87 Specific gravity 65 104 Specification 6 Stability 49 63 Stability ...

Page 127: ...111 Appendices W Weighing 22 23 WindowsDirect 24 93 Z Zero range 62 Zero tracking 52 ...

Page 128: ......

Page 129: ...SC fully automatic span calibration is not activated operator must carry out span calibration with the built in weight refer to the following upon blinking of the Weight Symbol Clock CAL is a practical alternative that performs regular automatic span calibration according to time setting UW series only up to 3 times a day refer to main manual 10 3 3 Span Calibration Procedure Calibration is requir...

Page 130: ...blinking Peak Holding On 49 blinking Interval Timer Time interval 50 mark blinking Add on Mode Formulation Mode 52 51 mark blinking Animal Weighing On On On 53 and blinking A blinking Menu Group 5 CAL key seven times from mass display kg kg kg 54 mg mg mg 55 pcs Number PC 5 5 7 ct Carat CT 58 mom Momme MO 59 d Solid Specific gravity measurement 60 d Liquid density measurement 61 Lb Pound 61a Oz Ou...

Reviews: