78
17. Troubleshooting
17.2 Error Display
∗
Contact your Shimadzu representative.
Error display
Description
Countermeasure
Trouble in weight loading mechanism
Check transportation screws.
The load on the pan is unstable at cal-
ibration.
Avoid wind and vibration.
The drift of zero point is large at cali-
bration.
Empty the pan.
The drift is large at the time of PCAL.
Use correct weight.
The drift is large at span calibration.
Use correct weight.
Calibration weight is wrong.
Use correct weight.
Failure in the balance
(Stops when this is displayed)
∗
Received command code is not cor-
rect.
Check delimiter etc.
Integer of the displayed unit exceeded
7 digits.
Decrease the load.
Failure in the balance.
∗
PCAL password error.
Check the password.
Attempted to set improper numeric
value.
Review the numeric value and decimal
places.
[CAL]
key was pressed during menu
lock.
Release menu lock.
Supply voltage is abnormal.
Check the supply voltage.