18 Error Messages
18.1 Error Messages
297
GC-2014
18.1.7
Operations errors
Value outside of range was entered
When the numeric value entered is out of the valid range, this message or valid setting
range is displayed. Enter a valid number.
File operation errors
These massages appear when an incorrect file operation was attempted.
Press another key, and continue operation.
Overflow of calculated pressure value
The carrier gas pressure is calculated from the linear velocity, flow rate, or the split ratio. The
value you have input is outside the set range. Change the conditions so that the pressure is
within the set range, and enter a new value.
Overflow of calculated flow rate value
The carrier gas total flow rate is calculated from the split ratio or the pressure. The value you
have input is outside the set range. Change the conditions and enter a new value.
Code
Message
CS
5001
Input parameter out of range
No
Code
Message
CS
5002
Invalid file no.
No
5003
This file is now used
No
Code
Message
CS
5006
CAR1 calc. prss out of range
No
5007
CAR2 calc. prss out of range
No
5008
CAR3 calc. prss out of range
No
Code
Message
CS
5009
CAR1 calc. flow out of range
No
5010
CAR2 calc. flow out of range
No
5011
CAR3 calc. flow out of range
No
Summary of Contents for GC-2014
Page 4: ...This page is intentionally left blank ...
Page 12: ...Contents viii GC 2014 This page is intentionally left blank ...
Page 28: ...1 Installation 1 3 Gas Supply Plumbing 16 GC 2014 This page is intentionally left blank ...
Page 78: ...5 Analysis 5 2 Notes for Analysis 66 GC 2014 This page is intentionally left blank ...
Page 90: ...6 Basic Key Operation 6 4 Getting Help 78 GC 2014 This page is intentionally left blank ...
Page 282: ...16 Special Functions 16 10 ROM Version No 270 GC 2014 This page is intentionally left blank ...
Page 318: ...306 GC 2014 This page is intentionally left blank ...