![Shimadzu GC-2014 Instruction Manual Download Page 308](http://html1.mh-extra.com/html/shimadzu/gc-2014/gc-2014_instruction-manual_1256914308.webp)
18 Error Messages
18.1 Error Messages
296
GC-2014
Detector ignition errors
These messages appear when the FID or FPD do not ignite within a certain time after the
key is pressed.
Even when one of these messages appears, hydrogen gas keeps flowing when a manual
flow controller is used. Shut off the hydrogen gas for safety then check the following items.
(1) The column is connected
(2) Hydrogen is supplied at proper flow rate
(3) Air is supplied at proper flow rate
(4) Filament of the igniter is intact
(5) The jet of the FID is not clogged
(6) Unused FID is not set to On
This message appears when the APC (used to control detector gas) for hydrogen or air is
not Ready at the time of ignition. Check that the gas supply pressure is stable and whether
or not gas is leaking. If there is no problem with the gas supply, the hardware has failed. The
system cannot be used in this condition. Turn off the system and contact your Shimadzu
representative.
18.1.6
Program errors
Set value change
This message appears when a set value was changed while the program is running. If the
parameter or event has not yet been executed, the new value is used for the analysis.
Time over errors
These messages appear when the program execution time exceeds the maximum allowable
value (9999.99 min). Change the program so that its total execution time does not exceed
“9999.99 min.” Although the program can be executed even after this error occurs, it is
discontinued at 9999.99 min.
Code
Message
CS
4203
DET#1 ignition failed
No
4204
DET#2 ignition failed
No
4205
DET#3 ignition failed
No
4206
DET#4 ignition failed
No
Code
Message
CS
4207
H
2
, AIR APC are not ready
No
Code
Message
CS
4301
Settings were changed
No
Code
Message
CS
4302
Program time is over
No
4303
Clean up time is over
No
4304
Pre-run prog. time is over
No
Summary of Contents for GC-2014
Page 4: ...This page is intentionally left blank ...
Page 12: ...Contents viii GC 2014 This page is intentionally left blank ...
Page 28: ...1 Installation 1 3 Gas Supply Plumbing 16 GC 2014 This page is intentionally left blank ...
Page 78: ...5 Analysis 5 2 Notes for Analysis 66 GC 2014 This page is intentionally left blank ...
Page 90: ...6 Basic Key Operation 6 4 Getting Help 78 GC 2014 This page is intentionally left blank ...
Page 282: ...16 Special Functions 16 10 ROM Version No 270 GC 2014 This page is intentionally left blank ...
Page 318: ...306 GC 2014 This page is intentionally left blank ...