![Shimadzu GC-2014 Instruction Manual Download Page 306](http://html1.mh-extra.com/html/shimadzu/gc-2014/gc-2014_instruction-manual_1256914306.webp)
18 Error Messages
18.1 Error Messages
294
GC-2014
AOC communication errors
One of these messages appears during a link failure or communication failure. (When a
communication error occurs, the link is automatically disconnected.)
When one of these messages appears, check the connection status, and reset the link.
18.1.5
Detector errors
Over current
Current of the detector is abnormal. If the resistance of the filament becomes abnormally
high when the TCD or the FTD is in use, the protection circuit is actuated to prevent damage
to the filament, and an alarm sounds.
If either error has occurred, turn off the system.
Possible reasons for the protection circuit to be activated are described below.
• The set current value exceeds the maximum operating current.
• Gas is not flowing.
• Gas is leaking.
• A significant amount of air is present in the flow line (for TCD).
Correct the error then turn on the system. If the system does not recover after several
resets, or the reason for the error cannot be located, contact your Shimadzu representative.
Code
Message
CS
4009
AOC command is invalid
No
4010
AOC data is out of range
No
4011
AOC time out
No
4012
AOC2 TRS error
No
4013
AOC link error
No
Code
Message
CS
4101
DET#1 TCD cell error
Yes
4102
DET#2 TCD cell error
Yes
4103
DET#3 TCD cell error
Yes
4104
DET#4 TCD cell error
Yes
4105
DET#1 FTD current error
Yes
4106
DET#2 FTD current error
Yes
4107
DET#3 FTD current error
Yes
4108
DET#4 FTD current error
Yes
Summary of Contents for GC-2014
Page 4: ...This page is intentionally left blank ...
Page 12: ...Contents viii GC 2014 This page is intentionally left blank ...
Page 28: ...1 Installation 1 3 Gas Supply Plumbing 16 GC 2014 This page is intentionally left blank ...
Page 78: ...5 Analysis 5 2 Notes for Analysis 66 GC 2014 This page is intentionally left blank ...
Page 90: ...6 Basic Key Operation 6 4 Getting Help 78 GC 2014 This page is intentionally left blank ...
Page 282: ...16 Special Functions 16 10 ROM Version No 270 GC 2014 This page is intentionally left blank ...
Page 318: ...306 GC 2014 This page is intentionally left blank ...