GC-2014
18 Error Messages
18.1
Error Messages
This system is programmed to self-diagnose certain errors. When an error is detected, an
alarm sounds and one of the error messages shown below is displayed. The detected errors
are recorded in the “Error Log”.
For errors of which the “CS” column in the table below shows Yes, a screen to confirm about
resetting them appears. When the confirmation screen appears, select one of the following
actions:
Codes shown in the tables below identify errors. Provide the code when calling service
representative about an error. These codes are also recorded in the error log.
18.1.1
System errors
Power supply failure
If the above error messages appear, the hardware has failed. The GC system cannot be
used in this condition. Turn off the system and contact your Shimadzu representative.
Room temperature sensor/atmospheric pressure sensor error
The room temperature or atmospheric pressure is out of the performance range. If this error
occurs even though the actual room temperature/atmospheric pressure is within the range,
the sensors may have failed. The system cannot be used in this condition. Turn off the
system and contact your Shimadzu representative.
Reset Error
Resets the system parameters, restoring the conditions prior to the error.
System control resumes. If the cause of the error has been corrected, the
error does not occur any more.
Ignore Error
Select this to keep the system as it is for a while for some reason, for
example, in order to remove the cause of the error.
Code
Message
CS
0001
DC 5 V is out of range
No
0002
DC 24 V is out of range
No
0003
DC -15 V is out of range
No
Code
Message
CS
0005
Room temperature is out of range
No
0006
Atm. pressure is out of range
No
Summary of Contents for GC-2014
Page 4: ...This page is intentionally left blank ...
Page 12: ...Contents viii GC 2014 This page is intentionally left blank ...
Page 28: ...1 Installation 1 3 Gas Supply Plumbing 16 GC 2014 This page is intentionally left blank ...
Page 78: ...5 Analysis 5 2 Notes for Analysis 66 GC 2014 This page is intentionally left blank ...
Page 90: ...6 Basic Key Operation 6 4 Getting Help 78 GC 2014 This page is intentionally left blank ...
Page 282: ...16 Special Functions 16 10 ROM Version No 270 GC 2014 This page is intentionally left blank ...
Page 318: ...306 GC 2014 This page is intentionally left blank ...