15 Optional Devices
15.1 Auto Injector Parameters
202
GC-2014
15.1.10
Auto-sampler carousel and other optional parameters
Select [Sampler] (PF menu) from the AOC main screen to display the sampler screen shown
in Fig. 15.1.4.
Specify the options installed on the AOC, such as sampler carousel.
When AOC1 and AOC2 are selected for the analytical line in “Line configuration” of the
[SET] key, both “USE OF THE SAMPLER” and “USE OF THE SUB AOC” are automatically
set to “Use”.
15.1.11
Parameter list
USE OF THE SAMPLER
Selection: Use/Not Use, Default: Not Use
USE OF THE SUB AOC
Selection: Use/Not Use, Default: Not Use(Set to “Not Use” because it is not used by GC-2014.)
BAR CODE READER
Selection: Use/Not Use, Default: Not Use
Select “Use” when the Bar Code reader is installed.
Fig. 15.1.4
Sampler carousel and other options setup