14 Diagnosis
14.2 Log Reading Menu
183
GC-2014
14.2.4
Analysis log
When “2. ANALYSIS LOG” is selected, the analysis log is displayed. This log records
whether analyses were allowed to finish and whether monitored valves exceeded their set
valves.
(1) Screen description
Select “2. LOG READING MENU” from the [DIAG] key main screen, then select “2.
ANALYSIS LOG” to display the screen shown in Fig. 14.2.3.
25
−
545 logs are saved. The number of saved logs depends on the size of each log. If
the number of logs exceeds the capacity, existing logs are deleted starting from the
oldest.
During a temperature program, if the rate of temperature increase is too great, the GC may not be
within the Ready range. When this occurs, the analysis log shows “Fail”.
If there are more items than can be displayed on one screen, scroll through the screen using the left
and right arrow keys.
(2) PF menu
Fig. 14.2.3
Analysis log screen
PF menu
Description
Reference
section
Clear Log
Deletes all displayed logs.
When [Clear Log] (PF menu) is pressed, the screen changes and the
message “Clear log with PF2” appears. Press [Clear Log] (PF menu) to
clear the log.
――
Prints the analysis log to a Chromatopac.
――
NOTE
NOTE
Summary of Contents for GC-2014
Page 4: ...This page is intentionally left blank ...
Page 12: ...Contents viii GC 2014 This page is intentionally left blank ...
Page 28: ...1 Installation 1 3 Gas Supply Plumbing 16 GC 2014 This page is intentionally left blank ...
Page 78: ...5 Analysis 5 2 Notes for Analysis 66 GC 2014 This page is intentionally left blank ...
Page 90: ...6 Basic Key Operation 6 4 Getting Help 78 GC 2014 This page is intentionally left blank ...
Page 282: ...16 Special Functions 16 10 ROM Version No 270 GC 2014 This page is intentionally left blank ...
Page 318: ...306 GC 2014 This page is intentionally left blank ...