Settings
Settings
Log-in Function
Log-in Function ON/OFF
This is a function for heightening security for the balance with the use of user IDs and passwords.
Setting the log-in function to [ON] will enable the administrator to control who is using the balance and
what they are using it for. Examples of using the log-in function’s ON/OFF settings are shown in the
following table.
Log-In Function: OFF
Log-In Function: ON
All registered users
are recognized as
administrators.
(Default setting)
Administrator
User 01
User 10
Administrator authorization assigns the following to each user:
• User ID, User Name, Password Registration
• Authorizes or denies the use of [Amend Settings], [External Output of
Measurement Values], [Use USB Memory], [Sensitivity Adjustment],
[Sensitivity Test]
1.
Call out the user settings
(1) Press [
MENU] while in the weight measurement mode, and then press [
LEFT].
(2) Select [
System settings], and then press [
OK].
(3) Select [User Settings], and then press [
OK].
2.
Set the log-in function to ON or OFF
(1) Select [Log-in Function], and then press [
OK].
(2) Select [on] or [off], and then press [
OK].
Set this to [on] if the log-in function is in use, and then
proceed onto Step 3.
3.
Set the management information for each user.
This sets the management information for each user.
^
[User Management (User Name Changes, Function Restrictions, etc.)] (P.74)
Up to
10 users can
be controlled