Removing Static Electricity
1.
Check to make sure that the [POWER]
indicator light for the ionizer is illuminated
2.
Close the glass draft shield
3.
Press [
ION]
The [RUN] lamp on the ionizer’s indicators will be illuminated, and ion emission will commence.
The time of ion emission is set at 10 seconds (default setting when shipped from the factory).
Ion emissions will stop automatically after the time set for ion emissions has elapsed.
Halting ion emissions
Press [
ION] during ion emission to halt emissions.
When measuring the net weight of flaks or other glass containers
Using the AP Holder (AP225W-AD, AP225W standard accessory) and ionizer together allows
for efficient elimination of static electricity from containers to allow for accurate measurement.
Changing Ion irradiation time
1.
Call out System settings
(1) Press [
MENU], and then .press [
LEFT].
(2) Select [ System settings], and then press [
OK].
(3) Select [System settings], and then press [
OK].
Simplified method of calling out the setting menu
It is possible to call out the setting menu in a simplified manner by pressing [
ION] for an
extended period.
2.
Set the time for ion emissions
(1) Select [Ion irradiation time], and then press [
OK].
(2) Set the Ion irradiation time between 1 - 300 seconds, and then press [
OK].
3.
Return to the weight measurement mode
Press [
POWER].