background image

PRECAUTION

The performance of the device may detoriate if any of the following conditions occur:
     Operating the device outside the stated t e mperature a nd h umidity r ange.
     Storing the device outside the   stated te mperature a nd h umidity r ange.
     Mechanical shock (for example, drop test) 

or degraded sensor.

     Patient’s temperature is below ambient temperature.
     Portable and mobile RF communications can affect the device.

DIGITAL THERMOMETER

CONTENTS

INSTRUCTION M

   ANUAL

One Shikon Digital Thermometer, One Instruction Manual, One Storage Case

Model:SK-033

SPECIFICATIONS

SYMBOL EXPLANATION

 Direct Current 

 Type BF  Applied Part

Batch Code

Manufacturer

Consult Accompanying Documents

Storage and Transportation 
Temp Limit: -4

℉ ~

131

 

-20

55

Measuring Range:

Ambient operating range:

90.0

-109.9

(32.0

℃-

42.9

)

±

0.2

 (

±

0.1

) during 95.9

~107.6

 (35.5

~42.0

at 64.4

~82.4

 (18

~28

) ambient operating range

Liquid crystal display, 3 1/2 digits 

One 1.5 V DC. button battery (size LR41or SR41, UCC 392) 
Approx. 200 hrs of continuous operation or One year with 3 
measurements per day
12.2cm x 1.9cm x 1.1cm (L x W x H) 
Approx. 10 grams including battery 

Temperature: 41

℉~

104

 (5

℃~

40

)

Relative humidity: 15%~95%RH
Atmospheric Pressure : 700hPa ~ 1060hPa

Type:

Digital Thermometer (Not Predictive)

Direct Mode

Accuracy:
Operating mode:                        

Display: 

Battery: 

Battery life: 

Dimension: 

Weight: 

PRODUCT  ILLUSTRATION

Battery Cap

On/Off

  Button

LCD

Probe

Memory:

For storing the last measured value

Do not use the thermometer in your ears. It is designed for oral, rectal, and armpit (axilla) 
readings only.
Do not place the thermometer battery near extreme heat as it may explode. Remove the 
battery from the device when not in operation for a long time.

The use of temperature readings for self-diagnosis is dangerous. Consult your doctor for the 
interpretation of result. Self-diagnosis may lead to worsening of existing disease conditions.
Do not attempt to use the thermometer when it is wet as it may show inaccurate result.
Do not bite the thermometer as it may lead to breakage and/or injury.
Do not attempt to disassemble the thermometer. Doing so may result in inaccurate readings.
After each use, disinfect the thermometer in case the device is used by more than one person.
Do not force the thermometer into the rectum. Failure to do so may lead to injury.
Do not use the thermometer orally after using it in the rectal area.
This device is not to be used orally by children less than 2 years old.
Store the device at temperatures between 5

℃~

40

(41

℉~

104

). 

  

SHIKON Digital Thermometer is designed to measure the human body temperature in 
regular mode orally, rectally or under the arm. It provides quick and accurate reading of 
the body temperature. The device is reusable for clinical or home use by people of all age 
group including children under 8 years old with adult supervision. To understand it’s 
function better and to obtain correct dependable results, please read the following 
instructions carefully before using the device. 

Temperature: -4

℉~

131

 (-20

~55

)

Storage and transportation
condition:  

Relative humidity: 15%~95%RH
Atmospheric Pressure : 700hPa ~ 1060hPa

Ingress Protection Rating:  IP 22

Classification:

Type BF

Choking Hazard:

  Thermometer cap and battery may be fatal if swallowed. Do not allow 

children to use this device without parental supervision.

Reviews: