10 |
P a g e
INSTALLATION
R
EQUIRED
A
MBIENT
C
ONDITIONS
These units are intended for use indoors, at room temperatures between
15
°
C and 30
°
C (59
°
F and 86
°
F)
,
at no greater than
80% Relative Humidity
(at 25
°
C / 77
°
F).
Operating these units outside of these conditions may adversely affect its incubator temperature stability
and effective operating range. For conditions outside of those listed above, please contact your SHEL LAB
distributor to explore other options suited to your laboratory or production environment.
R
EQUIRED
C
LEARANCES
These clearances are required to provide air flows for ventilation and cooling.
4 inches (102mm)
of clearance is required on the sides and back.
2 inches (51mm)
of headspace clearance between the top of the unit and any overhead partitions.
E
NVIRONMENTAL
D
ISRUPTION
S
OURCES
Consider proximate environmental factors that can affect the chamber temperature and atmospheric
integrity when selecting a location to install the unit:
•
Ovens, autoclaves, and any device that produces significant radiant heat
•
High-traffic areas
•
Direct sunlight
•
Heating and cooling ducts, or other sources of fast-moving air currents
Power Cord
2” (51mm)
2” (51mm)
4” (102mm)
4” (102mm)
4” (102mm)
SMI7 30” (762mm)
All Other SMIs 26” (660mm)
Door Swing
Summary of Contents for SMI11
Page 1: ...SMI INCUBATORS 100 120 Voltage Installation Operation Manual SMI2 SMI6 SMI7 SMI11 SMI12...
Page 2: ...2 P a g e Pictured on cover SMI6 SMI2 SMI7 SMI11 SMI12 2 Stacked SMI6s...
Page 14: ...14 P a g e INSTALLATION...
Page 16: ...16 P a g e GRAPHIC SYMBOLS...
Page 18: ...18 P a g e CONTROL PANEL OVERVIEW...
Page 32: ...32 P a g e USER MAINTENANCE...