42
6 Errors
6
Errors
If errors occur during regular operation, a corresponding mes-
sage appears on the display. The appropriate error number and
name are indicated.
Error
code
LCD
(error name)
Cause
Remedy
10
W ZERO POINT
ADJUSTMENT
An error occurred
during adjusting
the zero point
Check the flow rate.
Repeat adjustment.
See section 3.2.3.1 and 5
11
M ZERO POINT
ADJUSTMENT
An error occurred
during adjusting
the zero point
Check pressure
0.0 mbar.
Repeat adjustment.
See section 3.2.3.1 and 5
12
P ZERO POINT
ADJUSTMENT
An error occurred
during adjusting
the zero point
Check pressure
0.0 mbar.
Repeat adjustment.
See section 3.2.3.1 and 5
13
W SPAN
ADJUSTMENT
An error occurred
during adjusting
the flow rate
Check the flow rate.
Repeat adjustment.
See section 3.2.3.1 and 5
14
M SPAN
ADJUSTMENT
An error occurred
during adjusting
the pressure
Check pressure.
Repeat adjustment.
See section 3.2.3.1 and 5
15
P SPAN
ADJUSTMENT
An error occurred
during adjusting
the pressure
Check pressure.
Repeat adjustment.
See section 3.2.3.1 and 5
51
int. RAM
SERVICE
An error occurred
in the internal
RAM
Only the SEWERIN
service may perform
troubleshooting.
52
XFLASH
SERVICE
An error occurred
in the program/
data memory
Only the SEWERIN
service may perform
troubleshooting.
53
AD Con.
SERVICE
An error occurred
in the analogue
converter
Only the SEWERIN
service may perform
troubleshooting.
54
DISPLAY
SERVICE
A display error
occurred
Only the SEWERIN
service may perform
troubleshooting.