6 Maintenance
│
55
Fig. 40: Measuring mode
The
Device inspection due
sym-
bol is displayed in the status bar.
Measurement of the response time
The response time must be measured quarterly.
6.4.3
Variants of the guided device inspection
(standard and ECO)
The guided device inspection can be performed in two variants.
The variants differ in whether the indication accuracy is tested
and whether a bump test is performed.
Device inspection
Standard device inspection
−
Device condition
−
Pump
−
Zero point
−
Indication accuracy
ECO
device inspection
−
Device condition
−
Pump
−
Zero point
−
Bump test
Fig. 41: Variants of device inspection with associated subtests
The
ECO
device inspection is only available in the
Warning
ap-
plication.