If a fault occurs during operation, an error message will appear
on the screen.
Error messages are displayed in the order in which they occur.
Up to five errors can be displayed.
Error messages continue to be displayed until the error is cor-
rected.
Overview of possible error messages
Error
code
Error message on the
display
Error correction
8
No calibration
PPM sensor adjustment
Adjustment CxHy ppm
required
9
No calibration
IR sensor adjustment
Adjustment required (
adjust-
ment CxHy
,
adjustment CO2
,
adjustment gas mixture …
)
10
Adjustment failed
Test gas
Check test gas concentration
11
Zero point
Adjustment required
Adjustment required
52
XFLASH
SEWERIN Service
Error can only be corrected by
SEWERIN Service
59
Error unknown
SEWERIN Service
Error can only be corrected by
SEWERIN Service
62
IR sensor
Error can only be corrected by
SEWERIN Service
100
Pump error
Probe/filter
Check all filters, probes and hose
connections for porosity and dirt
200
I2C HOST – IR
SEWERIN Service
Error can only be corrected by
SEWERIN Service
201
I2C HOST – EC
SEWERIN Service
Error can only be corrected by
SEWERIN Service
202
I2C HOST – EX
SEWERIN Service
Error can only be corrected by
SEWERIN Service