56
│ 11 Faults and problems
Information
Information
Explanation/corrective action
Adjustment not possible.
Gas is missing.
A test gas required for the adjust-
ment is used up or not connected.
Therefore, the adjustment cannot be
carried out.
– Connect missing test gas.
– Check gas assignment.
Device inspection not
possible because no
device inspection is due.
Message appears in
Automatic
and
Timer
operating modes if all the
device inspections due on one day
have already been carried out.
–
Automatic
: Remove the device.
Do not reinsert it until the next
device inspection due date.
–
Timer
: The device can remain in
the test set. The device inspection
starts at the next specified time.
Device inspection
not possible or only
partly possible. Gas is
missing.
A test gas required for a device in-
spection is used up or not connect-
ed. The device inspection cannot
therefore be carried out or cannot be
carried out for all applications/sub-
tests that are due.
– Connect missing test gas.
– Check gas assignment.
Device inspection/
adjustment cancelled.
Check gas assignment!
The device has cancelled a device
inspection/adjustment because the
measurement values are not plau-
sible.
Free space in protocol
memory: …
Displays the remaining memory
slots for protocols in the test set.
Appears when less than 11 memory
slots are free (of max. 100).
– Create space in the protocol
memory using
GasCom
: Read
protocols >
Download
>
Delete
.