│ 11 Faults and problems
11
Faults and problems
11.1
Messages in the event of faults
If a fault occurs during operation, a message will appear on the
display.
Error messages with error code
Error
code
Errors
Explanation/corrective action
F081 System error:
Pressure control
– Contact SEWERIN Service.
F082 System error:
Pressure sensor
– Contact SEWERIN Service.
F084 Communication
error:
ATS – PM
Check contacts!
The data exchange between the
test set and the device is not
working. It is possible that the
device is not inserted correctly
and therefore has no electrical
contact with the test set.
– Remove the device from the
test set. Then reinsert the de-
vice.
F085 System error: Outlet
pressure too high
Excessive pressure at the gas
outlet of the device.
– Contact SEWERIN Service.
F089 Memory error:
XFlash
– Contact SEWERIN Service.
F090 File error:
Gas file missing
Update file!
The gas file is managed in
GasCom
(
Settings
>
Gases
). It
contains the assigned test gases.
–
Transfer the gas file from
GasCom
to the test set.
F092 File error:
User file faulty
Update file!
The user file is managed
in
GasCom
(
Tools
>
User
management
). It contains the
authorised users.
–
Transfer the user file from
GasCom
to the test set.