7 Device inspections
│
37
7.3
Features
7.3.1
All operating modes
●
The test set automatically remembers successful subtests for
use in subsequent device inspections. The requirement for this
is:
−
The device inspection or subtest was performed on the
same day.
●
If device inspections (or subtests thereof) are not technically
possible, a message will appear.
● Following certain subtests, the gas path is automatically
purged. If there is a waiting time associated with this,
Purge
will appear on the display. The device inspection will then con-
tinue automatically.
●
Device inspections are then automatically saved as a protocol.
−
Each device inspection is saved as a separate protocol.
−
If the device condition is also tested as part of a device in-
spection, the device inspection can be saved with a user
name.
1
In all other cases, the serial number of the test set is saved.
7.3.2
Manual operating mode: Special features and variants of
the device inspection
●
In
Manual
operating mode the device inspection must be per-
formed separately for each application.
−
Alternative: the test set will automatically perform all device
inspections that are due one after the other via the
All
menu
item.
●
In
Manual
operating mode, device inspections that are not due
can also be carried out if they are technically possible.
1
Setting using the
GasCom