6 Settings
│
33
●
Automatic
The device inspection starts as soon as a device is inserted
into the test set that is switched on.
All the device inspections that are due and are technically pos-
sible are carried out.
●
Timer
The device inspection starts at the specified time, provided
there is a device in the test set that is switched on.
All the device inspections that are due and are technically pos-
sible are carried out.
6.3.1.3 Device inspection mode
If a device inspection fails, an adjustment must be performed.
1
This adjustment can be started automatically by the test set or
manually.
●
Without adjustment
The adjustment must be started manually by the user.
●
With adjustment in case of error
An adjustment is started automatically following a failed device
inspection.
The test set performs the adjustment immediately following the
automatic subtests performed (section 7.4). The device inspec-
tion is repeated after the adjustment. The
Device condition
subtest is then performed.
In mode
With adjustment in case of error
, the corresponding
symbol (plus sign) is visible in the status area (fig. 25).
Fig. 25:
Manual
mode, device
inspection mode
With adjust-
ment in case of error
(symbol in
status area)
1
Does not apply if only the
Device condition
subtest failed but all other
subtests passed.