6.2.6  Each  PC  is  individually  checked  (visual  inspection  of  fittings,  helium  leak  test  and 

flow calibration). Sevenstar shall not be responsible for any damage caused by gas leakage 

or  the  use  of  a  dangerous  gas.  Users  are  responsible  for  following  the  safety  rules 

applicable to each gas they use. Improper use of a Sevenstar PC will void the warranty, and 

PC that have been damaged as a result of improper use will not be replaced by Sevenstar. 

6.2.7 Specific warranty requirements are as follows : 

A

Gas must be clean and particle-free, which means a filter must be fitted in the gas 

line upstream of the PC. 

B

Gas  must  comply  with  the  following  pressure  specifications:  Gas  pressure  must 

never exceed 3MPa. 

C

Electrical connection requirements are as follows: 

The system must be wired carefully: non-observance of the pin-out may irreversibly 

damage the electronic board inside the PC, in which case the warranty will be invalidated. 

D

Gas connections: The fittings must be handled carefully. Sevenstar guarantees that 

all fittings have been individually inspected and are scratch-free. 

E

Fitting  procedure:  The  fitting  procedure  set  out  in  the  manual  must  be  followed 

meticulously. Specifically, the purge procedure is very important if corrosive gases or toxic 

gases are used. 

F

The PC must not be dismounted: The PC warranty will be invalidated if the seal 

between the PC block and cover is torn. 

 

  6.3 Services 

 

Sevenstar  can  provide  services  like  start-up  service,  software  development,  gas 

system design, training, etc. 

Please visit www.mfcsevenstar.cn for more information

Summary of Contents for PC100-D PC

Page 1: ...REV 1 0 PC100 D PC User s Manual 2 0 2 0 0 9...

Page 2: ...STALLATION 5 2 Troubleshooting 17 2 1 General 7 SECTION 6 WARRANTY AND SERVICES 2 2 Unpacking 7 6 1 Guarantee of Sevenstar 18 2 3 Mechanical Installation 7 6 2 Product Warranty 18 2 3 1 General 7 6 3...

Page 3: ...ate store and distribute any part of this manual in mean of electric mechanical photocopy recording or other way without permission of Sevenstar This manual is not assured that there is no mistake and...

Page 4: ...out any authorized and make sure that the label or and seal of product is not removed when returning for reworking recalibration and maintenance b Please contact professional for technical service Do...

Page 5: ...ose valve completely to make sure that there is not error flow 1 4 General Pressure controller PC accurately measures and controls mass flow rates which is widely applied in the fields as semiconducto...

Page 6: ...gital monitoring PC100 D series have wide range of power supply available for 15 24VDC Customer Secondary Development of control and sample software is available through the open protocol PC100 D PC d...

Page 7: ...10 to 50 Mechanical interface 1 4 Compression Fitting 1 4VCR M Valve Type Normally Closed Leak Integrity External 1 x 10 10 scc sec He Leak Integrity Through closed valve 1 0 of orifice Full Scale Ni...

Page 8: ...centimeter per minute SLM standard liter per minute Standard state Temperature 0 C 273 15K Pressure 101325Pa 760mmHg At standard state the gas density will be constant The multiplication of density an...

Page 9: ...de is common plastic bag the inside is clean bag The outside bag should be removed in the entrance to the clean room In order to minimize the contamination the second clean bag should be removed in th...

Page 10: ...8 of 22 Figure 2 1 PC100 D...

Page 11: ...9 of 22 Figure 2 2 Product with Compression Fitting1 4 imensions...

Page 12: ...Product with VCR1 4 M dimensions Attention The height which is showing in figure 2 2 and figure 2 3 of 138 2 mm without electric connectors of cable It should be added around 50mm more after adding th...

Page 13: ...finger tight Scribe both nut and body in order to mark the position of the nut While holding the body with a wrench tighten the nut 1 8 turn past finger tight for 316L stainless steel and nickel gaske...

Page 14: ...ut Please note you should use two spanners to operating one spanner for locking the fitting stable and another one for revolving the nut Especially when you dismantle the tube you should operate by th...

Page 15: ...13 of 22 2 4 2 Connections The 9 pin male Sub D connector RB and RM are shown in figure2 7 Figure 2 7 9 pin male D sub connector...

Page 16: ...uld be completed 2 5 1 Check that tubing is leak free 2 5 2 Check the process sequence and proper function of all other gas components involved 2 5 3 Check the voltage of command signals and power sup...

Page 17: ...he PC in digital or analog mode For more information please refer to www mfcsevenstar cn Path Service Downloads Software download Communication Protocol 3 3 Valve Type The type of PC100 D valve is Nor...

Page 18: ...ust liquid and oil stain If necessary the filter should be added to gas system for purification If the outlet of PC is connected to liquid sources a One way valve should be added to avoid the liquid b...

Page 19: ...essure incorrect Check pressure condition Wrong control mode Change the control mode by the software Power failure Check Power and pin position Setpoint incorrect Check setpoint signal Contamination C...

Page 20: ...products are guaranteed against defects in materials and workmanship if used in accordance with specifications and not subject to physical damage contamination alteration or retrofit Warranty periods...

Page 21: ...ure must never exceed 3MPa C Electrical connection requirements are as follows The system must be wired carefully non observance of the pin out may irreversibly damage the electronic board inside the...

Page 22: ...on and unreasonable usage 3 Operating and storing in inappropriate or execrable circumstance 4 Usage of instrument beyond user s manual 5 Unauthorized change or replacement of product For example It i...

Page 23: ...Torr 1200M 1200mbar 0000X customized pressure please consult d for details e Control Modes U up a a Electronic connector RB DB9 B model RS485 anolog RM DB9 M model RS485 anolog f f Fittings CC Compres...

Page 24: ...of Valve Normally closed Orifice Full Scale 2SLM Pressure Full Scale 1200mbar Control Modes up Electronic connector DB9 B model RS485 anolog Fittings Compression Fitting 1 4 Seal Materials Metal For...

Page 25: ...o 368 Zhangjiang Road Pudong New District Shanghai Tel 86 21 63532370 Fax 86 21 63533265 Shenzhen Office Room 202 Section B No 1 Chuangjin No 125 Chuangye Erlu 28th Section Baoan District Shenzhen Zip...

Reviews: