29
higher). The scale will return to the normal weighing mode and the bar code
print mode may be accessed by pressing the
./rcl
and
keys.
4. Press the
setup/cal
key. Press
clear
if you do not wish to proceed.
5. Read the display. The scale now asks if you wish to print the tare weight.
6. Enter
1
and press the
setup/cal
key.
7. Read the display. The scale now asks if you wish to print the gross weight.
8. Enter
0
and press the
setup/cal
key.
9. Read the display. The scale now asks if you wish to print the APW function.
10. Enter
0
and press the
setup/cal
key.
11. Read the display. The scale now asks if you wish to print the accumulated
weight (or count).
12. Enter
0
and press the
setup/cal
key.
13. Read the display. The scale asks if you wish to print the quantity (count).
14. Enter
0
and press the
setup/cal
key.
Your scale is now setup to produce bar code labels with a Setra Auto Count 200
system. For more information on available fields to print, refer to the Auto Count 200
Operator's Manual. Be certain to set the scale’s baud rate to 2400.
FORMATTING BAUD RATE
Your Setra scale is capable of interfacing with a wide variety of computer equipment.
To set the baud rate (the rate at which your scale sends information to a computer),
use the following procedure:
1. Press the
—
key and enter
5555
.
2. Press the
setup/cal
key. The display will read “CAL br”.
3. Enter the baud rate. Choose either 300, 600, 1200, 2400, 4800, or 9600.
4. Press the
setup/cal
key.
5. REPEAT STEPS 1-4 TO ENSURE THIS BAUD RATE IS RETAINED EVEN IF
POWER IS REMOVED FROM THE SCALE.
Note:
A baud rate of 2400 is automatically assigned when none is formatted.