SERVOTOUGH LaserExact Monitor, Operator Manual 02950001A_1
SERVICE
PROGRAM
27
mode (0 (ok)/ 1 (zero)/ 2 (span)). Parameters can be added to or deleted from the
logging file without interrupting the logging process. The configuration of the
logging file may be modified (e.g. specify text qualifier, data separator, add
comment on top etc.).
The second possibility to log data is
Graph logging
. In this case the data are
logged into an ASCII file, which is compatible with data presentation programs.
The output format cannot be selected, the filename cannot be changed and <New
file at midnight> is always selected.
4.3.4
View error log
After pressing <View error log> the program downloads the log of the instrument’s
errors and warnings (see Section 5.4 for description of the errors) including the
date and time of their last activation and deactivation. The currently active errors
and warnings are marked with a cross (x), while the inactive errors are marked
with a dash (-). <Save error log> saves the current list of errors to a text file.
<Clear error log> clears all the errors and warnings from the instrument’s internal
memory.
The instrument also stores the log (System log) of internal status and self-test
information. This information is useful for advanced diagnostics and fault-finding
and downloaded to a text file by pressing <Save system log>. Older system
information may be downloaded by pressing <Rewind system log 100 lines> and
subsequently <Save system log>.
Summary of Contents for SERVOTOUGH LaserExact 2950
Page 2: ...This page intentionally blank...
Page 6: ...SERVOTOUGH LaserExact Monitor Operator Manual 02950001A_1 IV This page intentionally blank...
Page 69: ...SERVOTOUGH LaserExact Monitor Operator Manual 02950001A_1 APPENDIX 63...
Page 70: ...SERVOTOUGH LaserExact Monitor Operator Manual 02950001A_1 64 APPENDIX...
Page 71: ...SERVOTOUGH LaserExact Monitor Operator Manual 02950001A_1 APPENDIX 65...