Plasma Gas Analyzer
46
02001001A_9
5.8.4 System Gain
This value is used to adjust the last stage of amplification. If for any reason the same amount
of impurity gives less signal, you may adjust this valve to set the amplification. You may see it
as a span potentiometer.
N.B.: You must recalibrate the analyzer each time you change the system gain.
The smallest gain is 2 and the highest is 1000
Minimum: 2
Maximum: 1000
Resolution: 0.01
Figure 36 - SYSTEM GAIN
<<<SYSTEM GAIN>>>
Gain:3.00 Actual:3.00
Signal:87000
Sample flow(CC):75.0
F3:NEXT F4:RET
Detector
G
Analog to
Digital
converter
System
gain
Summary of Contents for SERVOPRO Plasma
Page 1: ...Part Number 02001001A Revision 9 Language UK English SERVOPRO Plasma OPERATOR MANUAL...
Page 2: ...This page intentionally blank...
Page 10: ...Plasma Gas Analyzer 8 02001001A_9 Figure 3 INTERNAL SCHEMATIC...
Page 16: ...Plasma Gas Analyzer 14 02001001A_9 Figure 4 DIMENSIONAL OUTLINE DRAWING...
Page 58: ...Plasma Gas Analyzer 56 02001001A_9 APPENDIX 1 AUTOMATIC CALIBRATION OPTION...
Page 66: ...Plasma Gas Analyzer 64 02001001A_9 Figure 49 SERVOPRO PLASMA WITH AUTOMATIC CALIBRATION...
Page 67: ...Plasma Gas Analyzer 65 02001001A_9 APPENDIX 2 TRACE NITROGEN IN HELIUM VERSION...
Page 70: ...Plasma Gas Analyzer 68 02001001A_9 APPENDIX 3 DUAL BACKGROUND ANALYZERS...
Page 76: ...Plasma Gas Analyzer 74 02001001A_9 APPENDIX 4 INTERFACE FOR EXTERNAL VALVE BOX...