Plasma Gas Analyzer
35
02001001A_9
5.5
Diagnosis
Pressing
"F3"
from the
MAIN MENU
will bring you to the
DIAGNOSTIC MENU
.
The diagnostic feature of the analyzer gives you the opportunity to verify the output of the
analogue to digital converter count value and check the 4 - 20 mA output.
If you have any options other than standard, you may also verify them (digital and/or fully
automatic calibration option). Please refer to the appropriate section of this manual.
Figure 23 - DIAGNOSTIC MAIN MENU
In the diagnostic menu the cell counts and flow counts are displayed. It is very useful to verify
stability and cell offset.
By selecting "
F2
" it is possible to see the error and alarm historic menu.
“F3”
brings you to the second diagnostic menu.
By pressing the
"F2"
key, it is possible to control the analogue output manually. By selecting
"F4: RET"
, the system displays the
DIAGNOSTIC MAIN MENU
again. The display looks like
the following figure.
Figure 24 - DIAGNOSTIC MAIN MENU
<<DIAGNOSTIC MENU>> F3:NEXT
Cell counts: 13120
Flow counts: 1666432 Temperature: 24.45
F2:Error and Alarm historic F4:RET
<<DIAGNOSTIC MENU>> F3:NEXT
F2:4-20 mA output verification
F4:RET
Summary of Contents for SERVOPRO Plasma
Page 1: ...Part Number 02001001A Revision 9 Language UK English SERVOPRO Plasma OPERATOR MANUAL...
Page 2: ...This page intentionally blank...
Page 10: ...Plasma Gas Analyzer 8 02001001A_9 Figure 3 INTERNAL SCHEMATIC...
Page 16: ...Plasma Gas Analyzer 14 02001001A_9 Figure 4 DIMENSIONAL OUTLINE DRAWING...
Page 58: ...Plasma Gas Analyzer 56 02001001A_9 APPENDIX 1 AUTOMATIC CALIBRATION OPTION...
Page 66: ...Plasma Gas Analyzer 64 02001001A_9 Figure 49 SERVOPRO PLASMA WITH AUTOMATIC CALIBRATION...
Page 67: ...Plasma Gas Analyzer 65 02001001A_9 APPENDIX 2 TRACE NITROGEN IN HELIUM VERSION...
Page 70: ...Plasma Gas Analyzer 68 02001001A_9 APPENDIX 3 DUAL BACKGROUND ANALYZERS...
Page 76: ...Plasma Gas Analyzer 74 02001001A_9 APPENDIX 4 INTERFACE FOR EXTERNAL VALVE BOX...