User Interface DF-760E
61
7.5.5.3 Pump Capacity Test Log
The pump capacity test can be used to determine the ultimate vacuum that the aspirator or
pump is capable of pulling. Proper flow through the analyzer is a direct result of the proper
balance between the pressure on the analyzer inlet and the vacuum at the analyzer outlet. If
either is incorrect, the flow will suffer and the moisture readings will not accurately reflect
the condition of the process gas. See page 70 for additional information on the pump test.
As the last step of the test, the system automatically puts an entry in the pump capacity test
log for future reference. See Review of this information can be useful in detecting a trend
in the pump condition which can result in a need to rebuild the pump.
Figure 40: Pump Capacity Test Log
7.5.6 Moisture Alarm Setup Menu
Figure 41: Moisture Alarm Setup SubMenu
The moisture analyzer includes a total of seven alarms. The four moisture concentration
alarms can be user controlled to activate up to four optional relays. High and low setpoints
as well as deadbands are user-set.
Summary of Contents for NanoTrace DF-760E
Page 1: ...NanoTrace Dual Moisture Oxygen Analyzer DF 760E Instruction Manual...
Page 4: ......
Page 38: ......
Page 44: ......
Page 122: ......
Page 128: ......
Page 138: ......
Page 142: ......
Page 143: ...Service DF 760E 139...
Page 144: ...140 DF 760E...
Page 148: ......
Page 166: ......
Page 168: ......
Page 171: ...DF 760E 167...
Page 174: ...170 DF 760E Appendix A User Menu Screens Page 88 Page 91 Page 93 Page 97 Page 98 Page 99...
Page 176: ...172 DF 760E Appendix A User Menu Screens Page 110 Page 111 Page 113 Page 114 Page 115 Page 116...