36 DF-760E
Options
Figure 8: Vacuum Pump Assembly
5.6.1 Installation of the Vacuum Pump
Mount the vacuum pump to the bracket
Connect the line from the moisture sample outlet to the needle control valve and
vacuum pump
Make the electrical power connection to the vacuum pump
5.6.1.1 Vacuum Pump Mounting
Mount the vacuum pump within 8 feet of the analyzer. Refer to Figure 9 and Figure 10 for
mounting hole and pump dimensions.
CAUTION
Be sure the pump outlet is at atmospheric pressure before starting.
The pump is not designed to start against any backpressure.
PLASTIC TUBING
SUPPLIED
TO MOISTURE
SENSOR OUTLET
POWER CORD
TO VACUUM PUMP
POWER CONNECTOR