07931001B revision 6
© Servomex Group Limited. 2020
71
Raw signal graph
To view the raw signal graph which shows the raw gas and ABurst signal and to
check that both signals are not above 125,000 counts.
A raw signal above the limit will raise an error status and
incorrect readings.
Excessive gas concentration for the path length used may
result in unacceptably high raw gas signal.
Figure 4-33: Raw signal graph example
When the gain adjustment and laser alignment have been optimized no further gain
adjustments should be required.
ABurst Typically
100000 counts
Gas signal
Manage Current Config
Phase/Gain Settings
Service Mode
Alignment LEDs
Phase/Gain Settings
Alignment Lock Level
Measurement Signal
Photodiode DC Level
Physical Setup
DC Gain 1
Detailed Setup
DC Gain 2
Transmission Level
Raw Signal
select
AC Amplifier Gain
ABurst
Adjust Burst Phase Cal
Ref TransCal