Installation location
Make sure that you mount the Laser 3 Plus on a structure that is able to
support its mass (refer to 2.6.1 for the transmitter unit and for the
receiver unit).
The environmental operating conditions for the Laser 3 Plus are listed in
Section 2.6.5.
Make sure that all floors and platforms are clear and free from
obstructions, and that the engineer has sufficient space to move freely
and change posture.
The Analyzer may fail if you use materials to connect the Laser 3 Plus to
the process that are not compatible with the process environment.
When you decide where to place the Laser 3 Plus in the process, we recommend a
minimum of 5 stack diameters of straight duct before and 2 stack diameters of
straight duct after the measurement point. Recommended distance from the process
wall, duct or pipe to the welded flange is a minimum of 100 mm. This will need to be
increased, or thermal break added, for process temperatures above 135°C
.
Analyzer placement
You should place the transmitter and receiver units so they are easily accessible. A
person should be able to stand in front of the transmitter to view the display and
access the alignment mounting to make adjustments. Cabling exiting the Analyzer
should also be considered.
Each unit requires at least 500 mm (19.7 ") of free space (measured from flange to
the stack and outwards).
Figure 7-1: Laser 3 Plus installation distances
Summary of Contents for 07931001B
Page 2: ...Servomex Group Limited 2020 This page intentionally blank...
Page 4: ...Servomex Group Limited 2020 This page intentionally blank...
Page 12: ...viii Servomex Group Limited 2020 07931001B revision 6 This page intentionally blank...
Page 14: ...x Servomex Group Limited 2020 07931001B revision 6 This page intentionally blank...