background image

 

SeQual Technologies Inc. 

 
 

ECLIPSE™  
Model 1000 

 
 

User Manual 

 

 

 
 

PN 6112 

Rev X5 

01/06 

 

 
 
 

REV. 

CO No. 

DESCRIPTION 

REL. DATE 

PREP. BY 

APPR. BY 

X1 

D017 

Initial Design Release 

11/16/01 

SWB 

PJJ 

X2 D120 

Update manual to remove 
facemasks 

11/22/04 WB 

WJB 

X3 D250 

Update to current design 

8/4/05 PJ 

PJJ 

X4 D319 

Update to current design 

12/23/05 PJJ 

PJJ 

X5 D324 

SCR001 – Pulse Mode 
breath detection 

1/5/06 WJB 

PSA 

 
 

File Name

 

 6112_X5 

Doc. Number 

6112 

 

Revision: 

X5 

Page: 

1 of 

35

 

 

 

Summary of Contents for ECLIPSE 1000

Page 1: ...BY X1 D017 Initial Design Release 11 16 01 SWB PJJ X2 D120 Update manual to remove facemasks 11 22 04 WB WJB X3 D250 Update to current design 8 4 05 PJ PJJ X4 D319 Update to current design 12 23 05 PJJ PJJ X5 D324 SCR001 Pulse Mode breath detection 1 5 06 WJB PSA File Name 6112_X5 Doc Number 6112 Revision X5 Page 1 of 35 ...

Page 2: ...EL 10 Description of the ECLIPSE 10 Control Panel Features 11 User Controls 11 Front Panel ICONS 11 System Status Indicators 12 RECOMMENDED OPERATING ENVIRONMENTS 13 The Proper Location 13 OPERATING INSTRUCTIONS 14 Before Operating 14 Powering your ECLIPSE System 14 Operating your ECLIPSE System 18 ALARM CONDITIONS AND TROUBLESHOOTING 21 ACCESSORIES 23 CLEANING CARE AND ROUTINE MAINTENANCE 25 TECH...

Page 3: ...tential risk exists if the operating instructions are not followed Attention or Caution Consult accompanying documents This symbol indicates important information that requires the user to take special precautions when certain conditions are present No Smoking or Naked Flames The use of oxygen therapy requires that special care be taken to reduce the risk of fire Materials that burn in air and eve...

Page 4: ...Flow Setting Increases the flow setting in 0 5 LPM increments each time the button is pushed Decrease Flow Setting Decreases the flow setting in 0 5 LPM increments each time the button is pushed Flow Mode Activates either Continuous Mode or Pulse Mode Operation External Power Present Indicator Indicates the presence of external power to operate the concentrator Power Cartridge Status Shows the amo...

Page 5: ...RATES SHOULD BE ADJUSTED ONLY UNDER THE ADVICE OF A PHYSICIAN WARNING FEDERAL LAW RESTRICTS THIS DEVICE TO SALE BY OR ON THE ORDER OF A PHYSICIAN CONTRAINDICATIONS WARNING THE ECLIPSE IS NOT INTENDED FOR LIFE SUPPORTING OR LIFE SUSTAINING APPLICATIONS NOR DOES IT PROVIDE ANY PATIENT MONITORING CAPABILITIES WARNING IN CERTAIN CIRCUMSTANCES THE USE OF NON PRESCRIBED OXYGEN CAN BE HAZARDOUS THIS DEVI...

Page 6: ... flow rate unless otherwise instructed by your physician WARNING Warning No Smoking or Naked Flames The use of oxygen therapy requires that special care be taken to reduce the risk of fire Oxygen enhances the combustion of flammable material For safety concerns all possible sources of ignition must be kept away from the oxygen system and preferably out of the room in which it is being used Smoking...

Page 7: ...erfill the humidifier if used Fill with water only to the level shown by the manufacturer When using ECLIPSE in a vehicle make sure it is properly secured or restrained WARNING DO NOT remove the cover no serviceable parts inside Qualified service personnel should only remove the cover WARNING DO NOT operate device or accessories on wet surfaces or in standing water and DO NOT submerse or expose to...

Page 8: ...xcessive heat WARNING DO NOT disassemble puncture or crush the Power Cartridge Power Cartridge electrolytes may be toxic if swallowed and can be harmful to skin and eyes Keep the Power Cartridge away from children WARNING Use of a damaged Power Cartridge may cause personal injury WARNING Exposing the Power Cartridge to water or other liquids may cause personal injury CAUTION DO NOT expose the Powe...

Page 9: ... TERM OXYGEN THERAPY IS NOW WIDELY ACCEPTED AS THE STANDARD TREATMENT FOR CHRONIC HYPOXIA CAUSED BY CHRONIC OBSTRUCTIVE PULMONARY DISEASE OXYGEN THERAPY HAS BECOME MORE VERSATILE IN RECENT YEARS DUE TO THE DEVELOPMENT OF OXYGEN CONCENTRATORS THE ECLIPSE MAKES IT EASIER FOR PHYSICIANS TO SELECT A MODE OF OXYGEN THERAPY THAT SUITS THE PATIENT S LIFESTYLE THE ECLIPSE MAY BE USED IN HOME INSTITIUTION ...

Page 10: ...onnections Oxygen Outlet Port Extension tubing or a cannula is attached to this port Air Inlet Filter Ambient air is drawn into the unit through the air inlet located on the top back of the unit This filter prevents dust and debris from entering the Eclipse and should be cleaned regularly Service Port This connection allows service personnel to diagnose and monitor Eclipse performance It is not fo...

Page 11: ...the control buttons and function for your ECLIPSE Oxygen System Front Panel Icons ON OFF Button and Indicator This button turns the system ON or OFF The Green Indicator is lit when the unit is ON Increase or Decrease Flow Rate Button Use these buttons to set the flow rate to your prescribed level Flow Mode Button and Indicator The ECLIPSE contains a button to toggle between Continuous Flow and Pul...

Page 12: ...ider Buzzer An audible alarm or buzzer is used to alert you of the condition of the unit either a warning or failure and to confirm a valid key press 2 0 Flow Rate Setting Indicator Green This is the main focus on your control panel Your dealer provider will set up your prescribed flow rates for your Continuous Flow Mode and your Pulse Flow Mode if pulse flow has been prescribed DO NOT change thes...

Page 13: ...ime for your ECLIPSE to acclimate to the recommend operating environment WARNING Operating your ECLIPSE outside the recommended operating environment may impact performance cause damage and will void the warranty CAUTION Locate the unit in a well ventilated room that provides adequate air flow and check if air is being prevented from entering or exiting the system by furniture draperies bedding cl...

Page 14: ...are not blocked and the air inlet filter is clean before using your system POWERING YOUR ECLIPSE SYSTEM The ECLIPSE is a lightweight portable oxygen system capable of being operated directly from three different power sources 1 an AC Power Supply 2 a DC Power Supply or 3 a rechargeable Power Cartridge The ECLIPSE charges the optional Power Cartridge when external power AC or DC is present and when...

Page 15: ...either 15 AMP or 20 AMP Consult your vehicle s Operator s Manual or look at your fuse box in the vehicle Set the switch on the DC Power Supply to the appropriate one Use the 10 AMP position on the DC Power Supply for a 15 AMP vehicle fuse and use the 15 AMP position for a 20 AMP or greater vehicle fuse Start your vehicle Insert the DC cord in the proper receptacle on the side of the unit Next inse...

Page 16: ...e DC Power Supply is connected Immediately disconnect the DC Power Supply from the motor vehicles accessory outlet if a low voltage alarm sounds 3 RECHARGEABLE POWER CARTRIDGE The ECLIPSE system can also be powered by a rechargeable Power Cartridge Installing the Power Cartridge Align the Power Cartridge so that it will slide into the Power Compartment Push the Power Cartridge into the Power Compa...

Page 17: ...low Mode LPM Pulse Flow Mode Setting 0 5 3 9 hours 1 0 3 5 hours 4 9 hours 2 0 2 1 hours 3 9 hours 3 0 1 3 hours 3 3 hours 4 0 2 8 hours 5 0 2 4 hours 6 0 2 2 hours Operating time will degrade with Power Cartridge use and age Store your Power Cartridge in a cool dry location This will help to assure the longevity of your Power Cartridge CAUTION DOT UN Regulations require the removal of the Power C...

Page 18: ...e AC Power Supply and plug in the AC Power Cord to completely charge the Power Cartridge This may require up to five hours of uninterrupted charging This completes the calibration of the Power Cartridge WARNING DO NOT tamper with disassemble crush or heat the Power Cartridge above 140 F 60 C The Power Cartridge may present a risk of fire or explosion and will void the warranty CAUTION Store the Po...

Page 19: ... buzzer will activate in the event of an internal fault or power outage external power with no Power Cartridge installed The red yellow and green all OK indicators will initially light up and then the red and yellow indicators sequentially go out when the oxygen purity is within specification After initially turning ON the unit allow up to five 5 minutes for the unit to reach its specified perform...

Page 20: ...ction the length of the tubing cannula must not exceed 7 feet 2 1m WARNING As with all conserving devices the ECLIPSE may not be able to detect some respiratory efforts Step 5 Adjust Flow Rate to Prescribed Level Using the Flow Rate Setting buttons adjust the flow rate to the prescribed setting supplied by your physician WARNING DO NOT change from prescribed flow rate setting without consulting wi...

Page 21: ...ersists contact Dealer Provider O2 Concentration 70 Off Off Blinking Beeps Continue Use Check or clean air intake filter If condition persists contact Dealer Provider Flow Rate Error Blocked Flow Off Blinking Off Beeps Check tubing and or humidifier for obstruction Check clean air intake filter If condition persists past 10 minutes contact dealer provider U NIT N EEDS A TTENTION No Inspiration Det...

Page 22: ...d or no external power is present Install charged Power Cartridge or attach external power No Oxygen Malfunction Contact your dealer provider Restriction in tubing or humidifier Repair or replace tubing or humidifier Low Oxygen Concentration Filter restricted Clean or replace air inlet filter Malfunction Contact your dealer provider Restriction in humidifier or tubing Repair or replace humidifier ...

Page 23: ...to the oxygen outlet port of the oxygen system and then to the humidifier Oxygen Tubing Salter Labs Series 2000 or equivalent Tubing Internal diameter of the tubing should be no less than 3 16 0 48 cm Connect the oxygen tubing to the outlet port of the humidifier or directly to the oxygen outlet port if you do not use a humidifier Connect the other end of the tube to the nasal cannula if not alrea...

Page 24: ...6104 Your AC Power Supply enables the ECLIPSE to be connected to a grounded AC electrical outlet with the country specific power cord Please contact your dealer provider for the country specific power cord DC Power Supply Model 6105 Your DC Power Supply enables the ECLIPSE to be operated in an automobile or other vehicle with an accessory outlet Contact your dealer provider to set up the DC Power ...

Page 25: ...eaning the air inlet filter is the most important activity you will perform to keep your oxygen system performing properly Check this air filter daily and clean it at least once a week 1 Remove the filter 2 Wash the filter in warm water using a mild detergent solution 3 Rinse the filter thoroughly and squeeze out the excess water 4 Allow the filter to air dry 5 Reinsert the filter in the cabinet C...

Page 26: ...quires special care to assure a longer life and higher performance The SeQual Power Cartridge is the only approved Power Cartridge to be used Once a month completely discharge your Power Cartridge and fully recharge to calibrate the Power Cartridge Use a damp cloth or sponge and mild detergent to clean the Power Cartridge case and the latch WARNING DO NOT tamper with or try to repair the Power Car...

Page 27: ...Power Supply 100 240 VAC 50 60 Hz 12 V 195 W Hrs 200 Watts Max 150 Watts Max Operating Temperature 50 F to 104 F 10 C to 40 C Operating Humidity 10 95 at 82 4 F 28 C dew point Sound 48 dBA at 3 0 LPM in Continuous Flow Mode 40 dBA at 3 0 LPM in Pulse Flow Mode Dimensions H x W x D 19 3 inches x 12 3 inches x 7 1 inches 49 0 cm x 31 2 cm x 18 0 cm Weight 17 4 pounds 7 9 kg with Power Cartridge 13 9...

Page 28: ...ven setting Delivery of the bolus is triggered by detection of the end of inspiration followed by the start of a subsequent inspiration Sensitivity is dealer provider adjustable Based on this inspiration trigger signal a pulse of oxygen is released to the patient by an electronically controlled flow valve There is no mixing of oxygen with room air Range of Operation Settings Increments of 1 6 time...

Page 29: ...ator is dangerous and can permanently damage the device and void the warranty WARNING Use No Oil or Grease A spontaneous and violent ignition may occur if oil grease or other petroleum substances come into contact with oxygen under pressure Keep these substances away from the oxygen system tubing and connections and any other oxygen source DO NOT use any petroleum based or other lubricants WARNING...

Page 30: ...airflow is not blocked This may impact performance WARNING DO NOT leave the ECLIPSE or DC Power Supply plugged into the vehicle without the engine running or attempt to start the vehicle while the DC Power Supply is connected to the vehicle This may drain the vehicle s battery WARNING If the vehicle power source vehicle accessory connector drops below 11 5 Volts brownout the ECLIPSE will revert to...

Page 31: ...y may result WARNING DO NOT disassemble the ECLIPSE or attempt to perform any routine maintenance other that what is described in the Patient Cleaning and Care section below Disassembling the oxygen system may create a hazard of electrical shock and void your warranty Contact your dealer provider for any maintenance required WARNING Unplug Power Cords AC or DC Power Supplies before cleaning the ex...

Page 32: ... air inlet filter is clean before using your system CAUTION The AC power cord used with the ECLIPSE Oxygen System must meet the requirements of the country where the product is used CAUTION Use with grounded 3 pin receptacle DO NOT remove ground pin from cord Use only with SeQual supplied cords CAUTION DO NOT connect the ECLIPSE to an extension cord or electrical outlet controlled by a switch CAUT...

Page 33: ...dequate flow to deliver oxygen the length of the tubing must not exceed 50 feet 15 2m when using Continuous Flow Mode and must not exceed 7 feet 2 1m when operating in Pulse Flow Mode for adequate breath detection CAUTION When operating the ECLIPSE on the cart DO NOT bang drop or bump the unit while in transport CAUTION The filter should be free of liquid water and dry to the touch before reinstal...

Page 34: ... with the instructions may cause harmful interference to other devices in the vicinity However there is no guarantee that interference will not occur in a particular installation If this equipment does cause harmful interference to other devices which can be determined by turning the equipment off and on the user is encouraged to try to correct the interference by one or more of the following meas...

Page 35: ...ctured by SeQual Technologies Inc 11436 Sorrento Valley Road San Diego CA 92121 USA 858 202 3100 800 826 4610 www sequal com European Union Authorized Agent MeddiQuest Limited Business Technology Centre Bessemer Drive Stevenage SG1 2DX United Kingdom Tel 44 1763 222505 Fax 44 1763 222049 Dealer provider Name and Address Physician Name and Address SeQual and ECLIPSE are registered marks of SeQual T...