Example of Connection with commercial Quantel USA Laser with Source and
Control Type ICE450
Configure the Quantel USA laser
In the Quantel Laser Control box, access the flash lamp menu and ensure that the line "flash
sync" is equal to "INT".
Fig. 30. Quantel Laser Control Box, from there it will configure the lamp and Q-
Switch to externally use the retarder
The following image shows the flash lamp setup menu for your Quantel laser. The lamp will be
triggered by the laser control itself, so it is correct to configure the lamp trigger menu or "Flash
sync" as "INT". The rest of the parameters of the laser lamp menu will depend on the conditions of
your experiment, during a set-up it is recommended to use the lowest possible lamp voltage.
Fig. 31. Laser flash lamp menu, with internal mode selected
Then access the configuration menu of the optical switch triggering or Q-Sw, to configure it in
external mode (EXT), since the Q-Switch will be externally controlled by the DG-3O1I-17. See the