3 Maintenance/Safety & Functionality Tests
HB-005615-j
16
3.1.6
Sensor Temperature Surveillance
Note: The ‘Sensor Temperature’ Icon in the Status Bar indicates the measured sensor
temperature (in °C).
Connect the sensor to the SDM and switch on the monitor. Verify that the sensor
temperature reaches the SET temperature within 2 to 3 minutes. Then immerse the
sensor into clean tap water of 45°C - 50°C and verify that the measured sensor
temperature increases (i.e. by observing the temperature indicator located in the
Status Bar of the SDM`s display). Verify that the red sensor LED switches off, a low
priority alarm starts to sound and that the message "Temp. limiter active" (or ‘Sensor
problem 38’, ‘Sensor problem 42’) displays.
If the message “Sensor Fault xx” was triggered you must restart the SDM to reset the
message. Do not use the sensor if the message “Sensor Fault xx” is not reset by power
cycling the SDM. Instead contact qualified service personnel or your local SenTec
representative.
Do not use the sensor if the test fails!
CAUTION: Ensure that the temperature of the water bath is
< 50°C. Temperatures above 50°C may damage the sensor.
3.1.7
Cleaning / Inspecting without Membrane
1.
Remove the sensor membrane using the membrane remover located at the
bottom of the Membrane Changer as shown in the figure below (note that the
Membrane Changer is shown upside down). Step
: Slide the SenTec TC Sensor
into the membrane remover with its membrane facing the bottom of the
Membrane Changer. Step
: Lift the SenTec TC Sensor up to remove the
membrane from the sensor body.
Figure 3
Removal of the Membrane from the TC Sensor
CAUTION: Do not touch any of the measuring units in the
center of the sensor surface after removing the sensor
membrane. Do not rub the sensor surface.