METIS M311 / M322 / H311 / H322 (17-pin)
Maintenance
50
10
Maintenance
CAUTION, switch off supply voltage.
Before beginning any maintenance, switch off power supply or pull connector on the py-
rometer to prevent injury by accidentally activated laser targeting light!
10.1 Cleaning
Clean the lens with a soft cloth and a little acetone. Use only high-purity acetone to avoid residues.
The objective lens is not
to clean with solvents that contain acid
immerse in water or other liquids to clean
10.2 Pyrometer Calibration
To obtain the measurement accuracy, we recommend to re-calibrate the pyrometer periodically (annual-
ly) with the help of calibration sources and optionally adjust at Sensortherm.
Fiber-optic pyrometers must always be calibrated together with the associated optics and the optical
fiber cable.
A recalibration is also necessary if (for fiber optic pyrometers) the optics OQ12 is replaced by OQ25
or vice versa and / or a new fiber optic cable should be used.
10.3 Trouble-Shooting
A device malfunction is indicated by a red power status LED. This may be a data transmission error or a
firmware problem. If the interface is not affected, the error can be read out via the
terminal box, see
). For interpretation of the answer please contact us so that we can find a solution
of the problem.
To further limit the possibilities, a service file can be created via the software
SensorTools
(see in
Device Settings and Configuration
) that can be sent to Sensortherm for further
support.