2
QUICK START INSTRUCTIONS
P
LEASE REFER TO THE REST OF THE INSTRUCTION MANUAL FOR MORE DETAILED INFORMATION
.
1. Find a suitable location that is clean, dry, and in a gas-free area.
Install the SMART-CAL 360 base on the bottom of the station
with the provided screws, as pictured. Place the assembly onto
a flat surface.
2. If SCAL-360
–D (desktop) or –N (network) software is to be
used, attach the appropriate USB or Ethernet cable(s). Refer
to instruction manual if utilizing Wi-Fi.
3. Attach the appropriate blade adapter to the power supply and plug it
into the power source and the back of the station. The station will power
on automatically, indicated by a blue status LED on the top of the unit.
4. Allow the station to complete the start-up process and reach the main
display. Press the CONTINUE button on the System Check screen to
acknowledge and move on to the Home Screen.
5. Prepare the calibration gas:
a. Attach the regulators to the gas bottles.
b. Attach the included hoses in the hose kit to the regulators, the unit
hose to the metal fitting on the back of the station, and the fresh air
hose(s) to the appropriate “Air” port connection(s) shown as shown
on-screen.
c. Attach the fittings on the hoses to the appropriate gas port connections on the back of the station as shown
on-screen. If the configuration shown does not match your needs, it can be re-configured in the Supervisor
Menu (refer to the instruction manual and contact SENSIT Technologies for the password).
d. If using positive pressure regulators with ON/OFF knobs, turn them on.
10. Power on a compatible SENSIT instrument to be calibrated or bump tested.
After the instrument completes the warmup process, put it into SMART-
CAL mode by pressing and holding the “A” button (left) for 1-2 seconds. If
performing a calibration, allow an additional 5-10 minutes of warmup time.
11. Place the instrument in the cradle and attach the instrument hose. Press
the CALIBRATION or BUMP button on the Home Screen of the station.
12. Allow the calibration or bump test process to complete. The result will be
given on-screen at the end, and the LED on the top of the unit will reflect
the result:
● WHITE – ready
● BLUE – busy / in progress
● YELLOW – check gas supply
● GREEN – calibration passed
● RED – calibration failed