SensComp MINI-S Installation And Operation Manual Download Page 1

General

 

This operation and installation manual provides 
general guidelines and suggestions to assist 
you in using the SensComp, Inc. MINI-S ultra-
sonic sensor module in many measurement 
applications.  For additional information, please 
contact a SensComp Applications Engineer at 
(734) 953-4783 between 9 AM and 5 PM EST. 

 

General Installation Procedures 

1.  Always mount the MINI-S in a suitable dry 

location.  The MINI-S is designed to be used 
in an indoor or protected environment only.  
The MINI-SE is suitable for harsher environ-
ments and higher humidity conditions. 
Excessive moisture on the circuit board (and 
the MINI-S transducer) will result in damage 
and improper operation, and will void all 
warranties. 

2.  Mount the MINI-S as far off the ground as 

practical. 

3.  Mount the MINI-S in a location where 

environmental interference sources are 
minimized (examples are EMI sources, air 
nozzles, excessive air turbulence, etc.). 

4.  Mount the MINI-S in a 1.575 inch diameter 

hole, using RTV silicone or edge clips to 
secure the sensor in place.  You can also 
use our Series 600 Housing unit, PID# 
619395, to house the MINI-S.   

5.  As supplied, the MINI-S has been calibrated 

and will function without further calibration.  

6.  The two (2) NPN open collector digital out-

puts need to be “pulled up” (through a resis-
tance and/or impedance) to an external 
positive power source of less than 40 VDC.  
(typically 4700 ohms when connected to an 
ex5 VDC power source for testing or 
for  interfacing to TTL Logic level  circuits).

  

The maximum output current is 600 mA. The 
Range Set LEDs do not require pull-up 
resistors, and can be used for setup 
calibration before connecting the MINI-S to 
external equipment. 

System Wiring Information

 

Pin 1 – Power Supply –  A +8 to +24 VDC 

regulated power source supplying 30 mA of 
current (2.0 Amperes during the 0.5 ms 
transmit pulse). 

Pin 2 – Power Supply Common (Ground) – 

Common Return for DC power supply, 
switched outputs, and clock signals. 

Pin 3 – External Trigger – Accepts TTL com-

patible logic level clock signals.   A low to 
high   (zero to +5 VDC) transition triggers 
the MINI-S 

Pin 4 – Trigger Enable – Allows the MINI-S to 

accept an external trigger signal. Enabled 
by connecting this pin (pin 4) to common 
(pin 2). 

Pin 5 – Clock Output – Delivers a TTL com-

patible Pulse Width Modulated (PWM) clock 
signal.  This signal goes high at the start of 
a cycle, and returns to a low state when the 
returning echo is received. 

Pin 6 – NPN Output 1 – This NPN open collector 

output turns on and off when a target is 
detected (on) or missing (off) as set by the 
Range Set 1 push button. 

Pin 7 – NPN Output 2 – This NPN open collector 

output turns on and off when a target is 
detected (on) or missing (off) as set by the 
Range Set 2 push button. 

 

Calibration Procedures 

1.  Apply DC power ( +8 to +24 VDC) to the 

MINI-S (connector header pins 1 and 2) 

2.  Allow five to ten minutes warm-up time for 

the MINI-S to reach operating tempera-
ture before calibrating the unit. 

3.  If desired for testing, connect the two 

NPN Open Collector Outputs to an 
external positive power supply through 
appropriate impedances.  Since they are 
open collector, they will always read zero 
volts until connected to a pull-up load 
(See General Installation Procedures – 
Step 6). 

4.  NPN Output 1 Adjustment (pin 6) 

  Place the target at the desired 

detection distance from the face of 
the MINI-S. 

  Depress and hold the “Range Set 1” 

push button switch. Wait for the 
“Range Window” LED to flash three 
times, followed by a “chirp” sound 
from the sensor before releasing.   

  The NPN output 1 should now be set:  

o

  LED 1 ON – NPN output is a 

Logic Low (0 VDC) when target 
is detected. 

o

  LED 1 OFF – NPN output is a 

Logic High (+ power source) 
when target is not detected. 

Reviews: