Chapter 7: IMS-4000 Sensors
125
Caution
: Removing a sensor for less than 60 seconds does not produce a trouble alarm.
Removing or unplugging a sensor for more than 60 seconds or plugging in a different
sensor at any time will cause a trouble alarm.
Host and Node Firmware Requirements
The Ultra Low Temperature Sensor requires the following firmware versions in the IMS-4000 Host
and updated ConsoleView software for your computer for sensor compatibility:
IMS-4000 Host
• Input board firmware chip version xx.xx.xx.106 or higher
NOTE
: Input board firmware is NOT flash upgradeable. A new firmware chip must be
installed. For information on obtaining a new firmware chip, please contact Sensaphone
Technical Support at 610.558.2700 or via e-mail at [email protected].
• ConsoleView version 3.0.15 or higher
• IMS4K OS version 3.00 or higher
• Voice version 2.12 or higher
*Host version information is viewable through the ConsoleView software on the Version Info
screen and the Help>About selection from the main menu.
IMS-4000 Node
• Firmware version 2.19 or higher (for 10Mbit/s nodes).
• All firmware versions (for 100Mbit/s nodes).
*Node version information is viewable through the ConsoleView software on the Node
Properties>Diagnostics screen.
Sensor Template
Input Name: Transducer
High Limit: 100
Low Limit: 0
Recognition Time: 3 Seconds
Reset Time: 0 Seconds
Data Logging: Active
Voice: None
Input Class: Other
Weight: 3 oz (85 grams)
Specifications
Temperature Range: -122 to 53°F (-86 to 12°C)
Humidity: 5–90% RH
Sensor type: sealed thermistor probe
Accuracy: +/- 3 degrees F
Connector: 7' cable w/RJ-45 plug
Housing: black plastic
Housing dimensions: 2.13'' l x 0.6'' d x 1.38'' h