Ideal for Screening for Lead, Mercury,

and Cadmium in Potable Water

(For Lead Soil Testing, see page 15)

Instruction Manual

Manufactured by:

Industrial Test Systems, Inc.

www.SENSAFE.com 

©2013 Industrial Test Systems, Inc. •  EXACT

®

 and LEADQuick

 are

 

trademarks of Industrial Test Systems, Inc., Rock Hill, SC, USA

 

Revision 02/20/13

For use with eXact

®

LEADQuick

 Photometer

For Your Safety:

Please read the entire manual 

before using the test kit.

Contains 50 Tests

U.S. Patent No. 7,333,194

South African Patent

# 2007/0628;

and other patent

applications including:

International Patent

Appln. No. PCT/US2005

/033985; and EU Patent

#1,725,864

This Instruction Manual does not include a 

test procedure for Lead in paint. Visit 

www.sensafe.com for details.

Parameter 

Page  MENU

Lead in Water 

PB2

Mercury in Water 

HG3

Cadmium in Water 

CD4

Spiked Recovery Method 

5

Tips for Best Accuracy 

6

Instrument Operation 

7

ReadySnap™ 1 Procedure  7
Kit Specifications 

8

Battery Installation 

9

Warranty (2 year)

 

9

Lead Recovery Samples 

10

Summary of Chemistry 

11

Lead Test Interferences 

12

MSDS Sheets 

13-14

Lead in Soil Method 

15

Kit Components 

16

Kit Number 486900-W (Lead and Mercury)

486900-W

Summary of Contents for eXact LEADQuick

Page 1: ...African Patent 2007 0628 and other patent applications including International Patent Appln No PCT US2005 033985 and EU Patent 1 725 864 This Instruction Manual does not include a test procedure for...

Page 2: ...into the CELL and immediately press READ This starts the 20 SECOND countdown timer During this time move the strip in a gentle back and forth motion Remove and discard the strip after 1 on the The di...

Page 3: ...y press READ This starts the 20 SECOND countdown timer During this time move the strip in a gentle back and forth motion Remove and discard the strip after 1 on the display disappears The display will...

Page 4: ...he 20 SECOND countdown timer During this time move the strip in a gentle back and forth motion Remove and discard the strip after 1 on the display disappears The display will flash and begin immediate...

Page 5: ...nt of interference within the detection limit of the test kit You can also consider calculating the actual lead in the sample see below as long as the Recovery is above a reasonable level such as 40 T...

Page 6: ...these strips may give unreliable results If conversion table s are supplied they have a unique revision number It is recommended that you visit www sensafe com to check for any updated revisions and...

Page 7: ...7 into cell and immediately push READ Move strip in a gentle back and forth motion approx 2 strokes Sec for 20 seconds Remove strip after 1 on the display disappears 8 Display will flash and automatic...

Page 8: ...ERO ON Button G Battery Cover with loop for lanyard H AAA Batteries x4 NOT INCLUDED eXact LEADQuick has been designed to offer the user a more Green and cost effective alternative to testing Instead o...

Page 9: ...L immediately after the test eXact LEADQuick Meter Messages eXact Photometer 2 Year Limited Warranty Registration of your eXact photometer must be received within 30 days from date of purchase to acti...

Page 10: ...52 95 Weaverville NC M 0 16 80 52 104 repeated 0 20 100 46 92 Anaheim CA M 0 19 95 53 106 repeated 0 21 105 47 94 Seattle WA M 0 16 80 45 90 repeated 0 16 80 42 84 Kilowna BC Canada M 0 19 95 51 102...

Page 11: ...est can tolerate From our lab and contract lab studies with LEADQuick we nd that over 90 of the typical potable municipally treated water samples experience no interference Before you start using the...

Page 12: ...he pH of the water needs to be adjusted to between pH 7 and 8 before testing begins Use 1 0N HCl or 1 0N NaOH to adjust the pH of your water sample Iron above 0 1 ppm can be present in well water samp...

Page 13: ...Oral LD50 17 000 mg kg rats Section 9 Other Information The above information is believed to be correct but does not purport to be all inclusive and shall be used ONLY as a guide Keep away from childr...

Page 14: ...ide Dispose of empty bottle and used test strip as normal trash Keep away from children and pets Store in a dry cool place Keep container tightly closed MSDS 4 Material Safety Data Sheet Section 1 Che...

Page 15: ...play disappears The display will ash and begin immediately counting up from 1 to 60 After the 60 seconds the cursor will move across the display informing you that it is about to measure the sample as...

Page 16: ...sured LeadQuick PB2 MENU ppb 0 4 6 6 10 18 63 110 209 322 0 5 10 10 7 14 20 60 107 214 321 428 438 0 4 10 10 10 20 67 110 221 331 451 0 4 8 8 10 19 65 110 215 326 5 444 5 Hach AAS Lead Std ppb Meter 1...