62580 SmartLog

X

3

™ 

 

 

Wrist Strap and Footwear Tester 

Automated ESD Record Keeping

  

 

Personal Grounding Verifier with Data Transmission 

Hardware & Software Setup Instructions

 

 

 

For sales and Service in 
North America contact:

 

ESD Systems.com 

www.esdsystems.com

 

[email protected]

 

Phone: 508-485-7390 

Fax: 508-480-0257 

Corporate Office: 

3651 Walnut Avenue, 

Chino, CA 91710 

Phone: (909) 627-8178 

Fax (909) 627-7449

 

For sales and service in Europe and Asia 

contact:

 

OK International - European Headquarters 

Eagle Close, Chandlers Ford, 

Hampshire, SO53 4NF U.K. England 

http://www.okintl.com/locations

 

Tel: +44 (0)23 8048 9100 

Fax: +44 (0)23 8048 9109 

 

Summary of Contents for SmartLog X3 62580

Page 1: ... in North America contact ESD Systems com www esdsystems com sales esdsystems com Phone 508 485 7390 Fax 508 480 0257 Corporate Office 3651 Walnut Avenue Chino CA 91710 Phone 909 627 8178 Fax 909 627 7449 For sales and service in Europe and Asia contact OK International European Headquarters Eagle Close Chandlers Ford Hampshire SO53 4NF U K England http www okintl com locations Tel 44 0 23 8048 91...

Page 2: ...Daisy Chained refer to next set of instructions for configuring clocks on a daisy chain Note Unit is not designed to work in high humidity environments where RH is above 65 RH Unpack and verify the following are included 1 SmartlogX3 on yellow mounting plate with tester and keypad 1 Dual foot plate 1 Stereo cable for dual foot plate 1 SmartlogX3 6 03 software 1 AC adapter 12VDC 500mA center pos 1 ...

Page 3: ...Advance button until the correct value appears on the display 2 From the end of the self diagnosis press the Menu button six times only you re giving a 20 sec limit to do so or clock will need to be reset by disconnecting power adapter to re enter set up menu Refer to Figure 1 for Button locations 3 Baud Rate should be set at 9600 factory default Press Enter for the next screen 4 Parity should be ...

Page 4: ...le provided inside box with unit Connect the cable from the right port RS 232 of the SmartlogX3 to the DB9 serial adapter connected to an open serial port on the PC used Cable is non inverted and flat type when laid across table top if user desires to make his her own cable Make sure cable is no longer than 50ft from the PC to the first SmartlogX3 2 SmartlogX3 to SmartlogX3 Data Connection These c...

Page 5: ...t Press CLEAR enter ID number and press ENTER to start test B Tester Wrist strap test range is 1Meg to 10Meg Ohms Footwear test range is 1Meg Ohms to 35Meg Ohms standard SmartlogX3 would have to be sent in to factory if test ranges is need to be set differently Tester test voltage is 12 0 VDC 0 40mA max for wrist strap and footwear C I R Barcode Reader reads the barcode from the right hand side ...

Page 6: ...nty should be shipped prepaid to the Semtronics factory Warranty replacements will take approximately two weeks If your unit is out of warranty Semtronics will quote repair charges necessary to bring your unit up to factory standards Call Customer Service at 909 627 8178 for proper shipping instructions and address Ship your unit freight prepaid WARRANTY EXCLUSIONS THE FOREGOING EXPRESS WARRANTY I...

Page 7: ...Testing Proper testing of the wrist strap includes the resistance of the groundable point on the end of the cord the cord itself the resistor the cord to cuff snap connector the resistance of the interface of the cuff the cuff wrist interface and the resistance of the person between the wrist and the hand that contacts the test electrode Per ESD Handbook TR 20 20 paragraph 5 3 2 2 2 Wrist Strap Gr...

Page 8: ...late could result in a false pass When using this setting special care must be taken to keep the footplate clean DIP switches 3 4 control the LOW limit for the footwear test selection as follows Switch 3 Switch 4 Low Limit Resistance ON OFF 100 Kilohms 1 X 105 OFF ON 1 Megohms 1 X 106 DEFAULT SETTING OFF OFF 10 Kilohms The Optional settings allow the use of custom resistance values when installed ...

Page 9: ... to three 3 seconds The resistance is checked from the touch plate to the corresponding foot plate for each foot and from the touch plate to the wrist strap connector jacks through the operator A wrist strap must be plugged into the appropriate jack before the touch plate is depressed if the wrist strap option is activated See wrist strap option selection on page 2 above The 62100 62102 can test b...

Page 10: ...2080 can be purchased for the convenient periodic The 62080 contains 14 standard resistors to check the following high and low limit ranges 1 10K low 2 10K safe 3 100K low 4 100K safe 5 1M low 6 1M safe 7 10M safe 8 10M high 9 35M safe 10 35M high 11 100M safe 12 100M high 13 1G safe 14 1G high TESTING PROCEDURE Wrist strap operation test 1 Insert the 62080 62100 62102 test plug into the dual wris...

Page 11: ...able connector on the 62100 62102 2 Select the appropriate fail low range on the rotary switch of the 62080 62100 62102 3 Press and hold the touch plate of the 62100 62102 until the test is completed The results of the test should be fail low on both feet 4 Select the appropriate safe low setting on the 62080 62100 62102 and repeat the test The results of the test should be pass on both feet 5 Sel...

Page 12: ...e Semtronics factory Warranty replacements will take approximately two weeks If your unit is out of warranty Semtronics will quote repair charges necessary to bring your unit up to factory standards Call Customer Service at 909 627 8178 for proper shipping instructions and address Ship your unit freight prepaid WARRANTY EXCLUSIONS THE FOREGOING EXPRESS WARRANTY IS MADE IN LIEU OF ALL OTHER PRODUCT...