12
Cleaning and Maintenance
EcoStar
Cleaning and Maintenance
WARNING
Always unplug the device from the electrical source and disconnect the respiratory circuit from the
device before cleaning.
CAUTION
Do not let water come into contact with the device.
Use appropriate materials for cleaning: do not use harsh detergents, abrasive sponges or brushes with
hard bristles.
Weekly
Air inlet filter
Remove the filter at the rear of the device.
Wash the filter with warm water and mild
detergent (for example 1 drop of
dishwashing liquid on the filter).
Rinse well to eliminate any trace of
detergent.
Dry the filter:
o
Press dry in a clean, absorbent cloth
o
Allow to dry completely away from
sunlight.
Once dry, reinstall the filter at the rear of
the device. Do not use a filter which is not
completely dry.
Monthly
Device
Clean the device exterior regularly by
using a damp cloth or paper towel
moistened with a little water and a
drop of gentle detergent.
Remove detergent residue by repeating
this step with a clean cloth or paper
towel, slightly moistened with just
water.
Wipe the entire device with a dry cloth
or paper towel.
Air inlet filter
Change the filter whenever it is damaged
or soiled.
WARNINGS
Do not use detergent sprays. Chemical
product residue could enter the air outlet,
the filter's foam or the device interior,
causing airway irritation.
Never use the device without making sure
that an air inlet filter is installed.
Note:
Please refer to the instructions for your mask, the heating humidifier and the respiratory circuit for more