Guardian Field Support Manual – Section 2
59
During the re-imaging process the status
light on the face of the ICS will
flash
bright
white
(with blue tint). When the process has
completed the status light will then transition
to a
steady bright green
indicating it has
completed successfully.
If the re-imaging has failed, the face of the
ICS will flash
bright red
with a repeated
‘chirp’ sound (no image for this status).
When the system has stopped flashing,
remove the USB Recovery Dongle.
Do not
pull the power as this may cause
corruption to the Software and a reimage
may be required.
Any interruption to the imaging process
prior to completion will corrupt the
recovery dongle. A new recovery dongle
will need to be made if this occurs.
The Controller will automatically restart.
During bootup an
Amber light
will be visible
for 90 secs. The processor will present the
Installation Wizard shortly after the Amber
light is turned off.
6.2 INSTALLATION WIZARD
Installation Wizard
The Installation Wizard will always start after you have:
- Used the USB Recovery Dongle for the first time.
- Used the USB Recovery Dongle to reset the system (re-flashing the system).
- Performed a Factory Reset.
You must have your Laptop setup to communicate to the Controller unit (refer to
).
Summary of Contents for G2-SY-CON2
Page 1: ...FIELD SUPPORT MANUAL Guardian Generation 2 Gen2 ...
Page 7: ...FIELD SUPPORT MANUAL Section 1 Introduction to Guardian Generation 2 ...
Page 29: ...1 FIELD SUPPORT MANUAL Section 2 Installation of Guardian Generation 2 ...
Page 100: ...Guardian Field Support Manual Section 2 72 ...