Secomam UVIKON XL Operating Manual Download Page 1

UVIKON XL 

UVIKON XS 

 

 
 
 

Operating manual 

 

 
 
 
 
 

Ref. 0M7717B 

 

 

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 

Summary of Contents for UVIKON XL

Page 1: ...UVIKON XL UVIKON XS Operating manual Ref 0M7717B...

Page 2: ...cking 17 6 1 2 Checking the serial number of the instrument 18 6 2 INSTALLING THE INSTRUMENT 19 6 2 1 Environmental conditions 19 6 3 INITIAL INSTALLATION 20 6 3 1 Connexion of the power supply 20 6 3...

Page 3: ...of this document does not confer any license to such patents The following names are AQUALABO ANALYSE s registered trademarks throughout the world SECOMAM UVIKON XL UVIKON XS This manual is updated p...

Page 4: ......

Page 5: ...ny breakdown due to the transporting of the equipment in packaging which is not its original packaging The lamps the cells and generally any item which appears in the accessories section on the price...

Page 6: ...are cannot be guaranteed exempt from defects or bugs TRADE SECRET AND PROPERTY RIGHTS This document is protected by a AQUALABO ANALYSE copyright c 2003 and the copying rights are explicitly reserved T...

Page 7: ...ment is carefully inspected before it is packed As soon as you receive your equipment check the condition of the packaging and if you notice any problems notify your carrier within 48 hours Then consu...

Page 8: ......

Page 9: ...y SECOMAM The user is permitted access only to the inside of the sample compartment Keep the sample compartment clean The Deuterium lamp used in this unit emits UV radiation Install the instrument in...

Page 10: ...g conditions are met a All electrical installations in the room or building in which the equipment is to be used must comply with regulations specified in the country where the equipment is to be used...

Page 11: ...NOTE Routine tests Routine test are performed on 100 of the produced equipment according to the requirements specified in Annex K of the Safety Standard EN 61010 1 A2 Dev UL CSA...

Page 12: ......

Page 13: ...capabilities The UVIKON Xx spectrophotometer is a double beam spectrophotometer which employs state of the art technology for collection presentation storage and retrieval of analysis data The UVIKON...

Page 14: ...Wavelength accuracy Holmium peaks 10nm min 0 05nm step 0 25 nm 0 3 nm Wavelength precision Holmium Oxide filter shoulder 0 025 nm 0 03 nm Baseline flatness 200 800nm 2nm 200nm min 1mAbs 1mAbs Stray li...

Page 15: ...halogen lamp on DEUTERIUM LAMP Green light indicates Deuterium lamp on or igniting POWER ON Green light indicates system power is on 5 3 2 Electronics Data processing is executed using single channel...

Page 16: ...UVIKON XL XS 16...

Page 17: ...other person help whenever lifting or moving the instrument Most of the accessories are contained in a separate enclosure Please make sure that you have removed these before storing the box do not dis...

Page 18: ...ON Software User Manual DNA RNA Purity Check Software User Manual 6 1 2 Checking the serial number of the instrument The plate showing the serial number is mounted to the rear right hand side of the i...

Page 19: ...ransient over voltages IEC 664 Pollution degree 2 CEI EN 61001 1 Room temperature between 15 C 59 F and 35 C 95 F In locations where the relative humidity is lower than 80 The UVIKON Xx contains many...

Page 20: ...of the equipment 6 3 2 Connection to the Personal Computer Use PC COM1 for the connection of the spectrophotometer serial cable and COM2 for external UVIKON XL accessories see section 6 Connect the se...

Page 21: ...on the spectrophotometer and the PC Launch Windows NT and press the corresponding LAB POWER icon in order to load the UVIKON spectrophotometer software If the communication is not active the sequence...

Page 22: ......

Page 23: ...inated and potentially hazardous Contaminated equipment and accessories must be decontaminated using the hospital decontamination and disinfecting procedures appropriate for the device Decontamination...

Page 24: ...ccessible from inside the sample compartment Keep the sample compartment clean To clean the instrument and the sample compartment use only a dry or water neutral detergent wet cloth do not use organic...

Page 25: ...ted magnetic bars set of 5 bars 8x3 mm 70 91 91671 The automatic sipper system is based on a peristaltic pump Different tube materials can be used Includes thermostatting fluid connection block and tu...

Page 26: ...K 2810 Q2 Stoppered microcells 2 mm inner width 10 mm path set of 2 70 K 2910 Q2 Open rectangular cells 10 mm path set of 2 70 K 2910 Q4 Open rectangular cells 10 mm path set of 4 70 K 3010 Q2 Stopper...

Page 27: ...hole in the cell holder This sensor can be used together with the magnetic stirrer 70 91 00580 Temperature sensor for cell content 70 91 00608 This sensor measures the sample temperature in a standar...

Page 28: ...rmation go to www aqualabo group com 2015 AQUALABO GROUP All rights reserved AQUALABO 90 rue du Professeur P Milliez 94506 Champigny sur Marne FRANCE Tel 33 0 1 55 09 10 10 Fax 33 0 1 55 09 10 39 info...