Operation • 53
17
-1
0-
05
-3
53
-0
02
g_
10
-2
02
0
B
• If inflation fails to deliver a measured value, the device automatically
switches to deflation. If necessary, the device re-pumps several
times during the deflation (re-pumping: Increasing cuff pressure by
approx. 50 mmHg followed by incremental releasing of that pres
-
sure). If no measured value is delivered even after several re-pump
-
ing steps, the procedure is canceled.
Starting a multiple measurement
►
Press the
button on the monitor.
The first part-measurement starts (in this case: 6 part-measurements).
Current cuff pressure is displayed.
The symbols for measurement procedure and measuring mode flash (in
this case: Inflation).
The part-measurement stops automatically as soon as valid blood
pressure values are detected.
The values for systolic/diastolic blood pressure
SYS/DIA
and mean
arterial pressure
MAP
are displayed.
The waiting time until the next part-measurement starts.
The next part-measurement starts automatically.
Once all part-measurements are present, mean blood pressure (mean
value) is displayed.
Disregarded part-measurements are displayed struck through (in this
case: Measurement 1).
You have the following options for continuing:
►
►
Measure bioimpedance
►
Stop the measurement procedure
NOTE
• Use the Start button to cancel the mean measurement and continue
or to cancel and re-start.
• If inflation does not deliver a measured value, the device automati
-
cally switches to deflation.
• If necessary, the device re-pumps several times during the deflation
(re-pumping: Increasing cuff pressure by approx. 50 mmHg fol
-
lowed by incremental releasing of that pressure). If no measured
value is delivered even after several re-pumping steps, the proce
-
dure is canceled.