3. Press the
buttons as appropriate:
►
Set display brightness
►
Set volume for warning and information sounds
►
Set volume for button sounds
NOTE
Amended settings can immediately be seen/heard each time a button
is pressed. You do not need to save or confirm them.
3.3 Calibrating the touchscreen display
1. Press the
Settings
tab.
2. Press the
Display and volume
button.
3. Press the
Calibrate touchscreen
button.
4. Confirm that you wish to proceed.
The calibration display appears:
5. Press the
symbol.
The symbol changes position.